Validation of the Reichert® Tono-Vera® Vet rebound tonometer in normal ex vivo canine eyes.

Kapeller Lydia E, Buckman Phillip N, Wang Sichao, Komáromy András M

СтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine the accuracy of the Reichert® Tono-Vera® Vet rebound tonometer for canine intraocular pressure (IOP) measurement.

ANIMALS STUDIED: Five normal canine ex vivo globes.

PROCEDURES: The anterior chambers of five freshly enucleated normal canine eyes were cannulated and connected to a reservoir of Plasma-Lyte A and a manometer. Starting at a manometric IOP of 5 mmHg, the pressure was progressively increased to 80 mmHg by raising the reservoir. Triplicate IOP measurements were taken with the Tono-Vera® Vet from the central cornea using the dog setting and compared to the manometric pressure by linear regression analysis and Bland-Altman plots.

RESULTS

There was a strong positive linear regression trend when comparing central corneal Tono-Vera® Vet IOPs to manometric pressures (r2 = .99) with solid agreement between the two methods. Compared to manometric IOPs, the Tono-Vera® Vet underestimated IOPs at higher pressures ≥70 mmHg.

CONCLUSIONS

Measurement of IOPs from the central cornea with the Tono-Vera® Vet provided accurate results over a large range in normal canine globes compared to direct manometry. The mild to moderate underestimation of IOPs at high pressures was not considered clinically relevant.

Для цитирования: Kapeller Lydia E, Buckman Phillip N, Wang Sichao, Komáromy András M. Validation of the Reichert® Tono-Vera® Vet rebound tonometer in normal ex vivo canine eyes.. Veterinary ophthalmology. 2024. DOI: 10.1111/vop.13213.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.