DOI 10.1016/s0002-9394(98)00228-1
A clinical comparison of transscleral cyclophotocoagulation with neodymium: YAG and semiconductor diode lasers.
Аннотация
PURPOSE: To compare the efficacy of transscleral cyclophotocoagulation using a neodymium: YAG (Nd:YAG) or semiconductor diode laser in controlling intraocular pressure in patients with refractory glaucoma.
METHODS
In a prospective study, 95 eyes of 91 patients with refractory glaucoma randomly received Nd:YAG or diode cyclophotocoagulation. Patients were followed for a mean of 10.4 months (10.42 +/- 3.16, mean +/- SD). We compared available data preoperatively and at 1 week, 1 month, 6 months, and 12 months postoperatively. Data analyzed were corrected visual acuity, intraocular pressure, and the type of glaucoma.
RESULTS
There was a statistically significant decrease in intraocular pressure after both Nd:YAG and diode cyclophotocoagulation at each time period. However, there were no significant differences in postoperative intraocular pressure or visual acuity change between Nd:YAG and diode procedures.
CONCLUSIONS
Compared with the Nd:YAG laser for transscleral cyclophotocoagulation, the diode laser has technological advantages including portability, durability, and smaller size, while providing equivalent postoperative intraocular pressure and visual acuity change.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.