DOI 10.1080/08037051.2018.1463817
Renal nerve stimulation identifies aorticorenal innervation and prevents inadvertent ablation of vagal nerves during renal denervation.
Аннотация
PURPOSE: Recently we reported the use of renal nerve stimulation (RNS) during renal denervation (RDN) procedures. RNS induced changes in blood pressure (BP) and heart rate are not fully delineated yet. We hypothesized that electrical stimulation of the sympathetic nerve tissue in the renal artery would lead to an increase in BP and vagal stimulation would cause a decrease in BP. We report the different patterns of BP and heart rate responses elicited by RNS prior to RDN.
METHODS
35 patients with drug-resistant hypertension were included. RNS was performed under general anesthesia at four sites in the right and left renal arteries, both before and immediately after RDN. RNS-induced BP and heart rate changes were monitored.
RESULTS
A total of 289 RNS sites in 35 patients were analyzed. An increase in systolic BP of >10 mmHg was regarded as a positive BP response to RNS. This pattern of response was observed in 180 sites (62%). 86 RNS sites (30%) showed an indifferent response with BP changes ≤10 mmHg. At 13 sites (4.5%) RNS elicited a decrease in BP up to -8 mmHg. However, 10 RNS sites (3.5%) showed a pronounced vagal response with hypotension and sinus cycle lengths ranging between 4224-10272 milliseconds. These sites were distributed among two patients.
CONCLUSION
RNS identified sympathetic and parasympathetic nerve tissue in the renal arteries. RNS can be potentially used to map nerve bundles and guide selective ablation of sympathetic nerve fibers and prevent inadvertent ablation of parasympathetic nerve tissue during RDN.