Accuracy of Portable Ultrasound Measurement of the Bladder Volume Using the Manual Fanning Scan and Traditional Motorized Scan Methods.

Jeong Hyun Ju, Oh Seung-June

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the accuracy of portable ultrasound bladder scanner (PUBS) for measuring bladder volume (BV) using manual fanning scan (MFS) and traditional motorized scan (TMS) methods.

METHODS

Patients with lower urinary tract symptoms who underwent a urodynamic study at the urology outpatient clinic from October to December 2024 were prospectively enrolled. BV was measured using TMS-type (TMS-BV) and MFS-type (MFS-BV) PUBS prior to free uroflowmetry. True BV (T-BV) was defined as the sum of the voided volume and the postvoid residual volume drained through the urethral catheter after voiding. TMS-BV and MFS-BV were compared and analyzed against T-BV.

RESULTS

Data from 39 patients (30 males and 9 females; mean age, 65.3±14.8 years) were analyzed. TMS-BV (230.9±157.9 mL) and MFS-BV (222.6±154.2 mL) did not differ significantly from T-BV (230.6±156.6 mL) (P>0.05). Comparison of the percentage of difference in volume between TMS-BV and MFS-BV across 100-mL BV ranges revealed no significant difference (P>0.05). The mean difference between TMS-BV and T-BV (0.2±50.2) did not differ significantly from the mean difference between MFS-BV and T-BV (-8.0±69.1) (P>0.05). The correlation between TMS-BV and T-BV (R2=0.90) was stronger than that between MFS-BV and T-BV (R2=0.81).

CONCLUSION

The correlation between MFS-type PUBS and T-BV was lower than that between TMS-type PUBS and T-BV. However, the measurement accuracy of BV was similar using both methods.

Для цитирования: Jeong Hyun Ju, Oh Seung-June. Accuracy of Portable Ultrasound Measurement of the Bladder Volume Using the Manual Fanning Scan and Traditional Motorized Scan Methods.. International neurourology journal. 2025. DOI: 10.5213/inj.2550190.095.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.