Efficacy and Safety of Ultrasonic Longitudinal-Axis Vibration for the Reduction of Ureteral Access Sheath Insertion Force: A Randomized Controlled Trial in a Porcine Model.
Аннотация
PURPOSE: Excessive bulking force during ureteral access sheath (UAS) placement may induce injury. The sliding friction between surfaces can be reduced with the application of ultrasonic vibration. We investigated the efficacy and safety of an ultrasonic vibration transducing device for reducing the maximal ureteral access sheath insertion force (UASIF).
MATERIALS AND
METHODS
A device was developed for transducing ultrasonic longitudinal-axis vibration onto the UAS at an adjustable amplitude and frequency while measuring the degree of UASIF. In the pilot study, six porcine models were used to investigate the optimal amplitude and frequency of vibration and to calculate sample size. Twelve porcine models were utilized in a randomized controlled trial. Resected ureters were pathologically evaluated for ureteral injury.
RESULTS
The transduction of ultrasonic vibration at an amplitude of 0.04 g and a frequency of 18,000 Hz resulted in a maximal UASIF reduction of 36.4% (interquartile range 32.7-43.1). Maximal UASIF tended to decrease with increasing vibration frequency. No significant differences in UASIF reductions were observed according to amplitude. In the randomized controlled trial, the maximal UASIF reduction was 37.0% (interquartile range 21.4-44.2). Grade II injury was pathologically diagnosed in 8.3% (1/12) of the ureters in both groups.
CONCLUSIONS
The transduction of ultrasonic longitudinal-axis vibration onto the UAS reduces maximal UASIF and does not harm the ureter. Reducing the velocity of sheath insertion may further reduce maximal UASIF.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.