MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00270-021-02915-0

Медицинская библиотека · Cardiovascular and interventional radiology · 2021

Ultrasound-Guided Percutaneous Laser Ablation of the Thyroid Gland in a Swine Model: Comparison of Ablation Parameters and Ablation Zone Dimensions.

Ridouani Fourat, Tuttle R Michael, Ghosn Mario, Li Duan, Wong Richard J, Fagin James A, Monette Sebastien, Solomon Stephen B, Camacho Juan C

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

To compare laser ablation (LA) zone dimensions at two predetermined energy parameters to cover a theoretical 10 mm zone + 2 mm margin in a thyroid swine model.

METHODS

Approval of the Institutional Animal Care and Use Committee was obtained. After hydrodissection, an ultrasound-guided LA (Elesta Echolaser X4 with Orblaze technology, 1064 nm) was performed in the periphery of the thyroid in 10 swine. Two cohorts were established to ablate a region of 10mm diameter with 2mm margin based on manufacturer's ex vivo data (n= 5 at 3W/1400J and n= 5 at 3W/1800J). The ablation zone was measured on contrast-enhanced computed tomography (CT) and compared to the pathological specimen. Euthanasia was performed 48 hours following ablation.

RESULTS

All ablations in the 3W/1800J group achieved a diameter of 12 mm ± 1 mm in three dimensions. In the 3W/1400J group, 1 ablation reached 12 mm ± 1 mm in 2 dimensions and 4 ablations reached this size in one dimension. Maximum diameter was higher in the 3W/1800J compared to the 3W/1400J group, both on histology (1.46 cm ± 0.05 vs. 1.1 cm ± 0.0, p< 0.01) and CT (1.52 cm ± 0.04 vs. 1.18 cm ± 0.04, p< 0.01). Similar results were obtained regarding volumes, both on histology (1.12 mL ± 0.13 vs. 0.57 mL ± 0.06, p< 0.01) and CT (1.24 mL ± 0.13 vs. 0.59 mL ± 0.07, p< 0.01). Histology showed coagulation necrosis and correlated well with CT measurements.

CONCLUSION

Optimal parameters to obtain a LA zone of 10 mm with 2 mm margin utilizing a single needle are 3W/1800J.

Для цитирования: Ridouani Fourat, Tuttle R Michael, Ghosn Mario, Li Duan, Wong Richard J, Fagin James A, Monette Sebastien, Solomon Stephen B, Camacho Juan C. Ultrasound-Guided Percutaneous Laser Ablation of the Thyroid Gland in a Swine Model: Comparison of Ablation Parameters and Ablation Zone Dimensions.. Cardiovascular and interventional radiology. 2021. DOI: 10.1007/s00270-021-02915-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.