DOI 10.1556/1886.2018.00020

Медицинская библиотека · European journal of microbiology & immunology · 2018

Poor Diagnostic Performance of a Species-Specific Loop-Mediated Isothermal Amplification (LAMP) Platform for Malaria.

Kollenda Hans, Hagen Ralf Matthias, Hanke Miriam, Rojak Sandra, Hinz Rebecca, Wassill Lars, Poppert Sven, Tannich Egbert, Frickmann Hagen

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The objective of this study was to assess an in-house loop-mediated isothermal amplification (LAMP) platform for malaria parasite detection and identification on species level.

METHODS

LAMP primers specific for the human Plasmodium spp., namely, P. falciparum, P. vivax, P. ovale, P. malariae, and P. knowlesi, as well as genus-specific primers, were tested against a composite gold standard comprising microscopy from thick and thin blood films, commercial genus-specific Meridian illumigene Malaria LAMP, in-house real-time polymerase chain reaction (PCR), and commercial fast-track diagnostics (FTD) Malaria differentiation PCR.

RESULTS

Of the 523 blood samples analyzed, the composite gold standard indicated 243 Plasmodium-species-DNA-containing samples (46.5%). Sensitivity and specificity of the analyzed genus- and species-specific LAMP primers were 71.0%-100.0% and 90.8%-100.0%, respectively. The influence of parasitemia was best documented for P. falciparum-specific LAMP with sensitivity values of 35.5% (22/62) for microscopically negative samples containing P. falciparum DNA, 50% (19/38) for parasitemia ≤50/μL, 84% (21/25) for parasitemia ≤500/μL, and 100% (92/92) for parasitemia >500/μL.

CONCLUSIONS

In our hands, performance characteristics of species-specific in-house LAMP for malaria lack reliability required for diagnostic laboratories. The use of the easy-to-apply technique for surveillance purposes may be considered.

Для цитирования: Kollenda Hans, Hagen Ralf Matthias, Hanke Miriam, Rojak Sandra, Hinz Rebecca, Wassill Lars, Poppert Sven, Tannich Egbert, Frickmann Hagen. Poor Diagnostic Performance of a Species-Specific Loop-Mediated Isothermal Amplification (LAMP) Platform for Malaria.. European journal of microbiology & immunology. 2018. DOI: 10.1556/1886.2018.00020.