DOI 10.1097/iop.0000000000000076
Conjunctivodacryocystorhinostomy and lacrimal sinus diversion via sinopsys surgical lacrimal stent: cadaveric proof of concept.
Аннотация
PURPOSE: To determine the viability of lacrimal sinus diversion by using a novel Sinopsys Surgical Lacrimal Stent.
METHODS
Two disarticulated cadaveric heads were used in the laboratory. First, bilateral conjunctivodacryocystorhinostomies (CDCR) were performed by using a traditional Pyrex Jones tube and the Sinopsys Surgical Lacrimal Stent. The fluiditics were then evaluated by using contrast agent and fluroscopy. Subsequently, conjunctivoethmoidectomies (CE) and conjunctivomaxillectomies (CM) were performed by using the Sinopsys Surgical Lacrimal Stent, of which the fluiditics were also studied.
RESULTS
The conjunctivodacryocystorhinostomy (CDCR) placement experience of the Jones tube and the Sinopsys Surgical Lacrimal Stent were similar. The CDCR fluiditics as measured by using contrast agent and fluoroscopy were identical with each draining 0.5 mL over 5 to 10 seconds. Placement of the Sinopsys Surgical Lacrimal Stent in the 4 ethmoid (CE) and 4 maxillary (CM) sinuses was similar to the CDCR experience. Fluiditics of lacrimal sinus diversion were similar to CDCR with drainage of 0.5 mL over 5 seconds.
CONCLUSIONS
The Sinopsys Surgical Lacrimal Stent has flow characteristics that are similar to a Pyrex Jones tube. Drainage in the nose via a CDCR procedure is similar to drainage in the ethmoid and maxillary sinuses via a CE and CM, respectively.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.