Qi X Sharon, Yang Lisa, Lee Percy, O'Connell Dylan, Chu Fang-I, Steinberg Michael L, Low Daniel A
Аннотация
PURPOSE: This study aimed to evaluate the potential of megavoltage-topogram (MV-topogram)-based alignment as an alternative to megavoltage computed tomography (MVCT) in reducing setup time and imaging dose for patients with malignant pleural mesothelioma who are receiving TomoTherapy.
METHODS AND MATERIALS: Twelve patients were enrolled in an ongoing institutional review board approved clinical trial at our institute. Patients were set up with a clinical protocol using red lasers. Anteroposterior (AP) and lateral (LAT) MV-topograms were acquired using gantry angles of 0°/90° with a 1 mm collimator opening, all multileaf collimator leaves open, a couch speed of 4 cm/s, and a 12.5-second scanning time. Routine MVCT scans were performed immediately afterward. The MV-topograms were reconstructed and enhanced using contrast-limited adaptive histogram equalization. Anteroposterior and LAT kilovoltage digital reconstructed topogram images were reconstructed based on TomoTherapy geometry from computed tomography simulation scans. Registrations between MV-topograms and kilovoltage-digital reconstructed topogram images were performed manually, and patients' daily shifts were recorded. Results were compared against the corresponding daily MVCT shifts. MV-topogram and MVCT doses were measured and recorded using an ion chamber on a cheese phantom with depths between 1 and 14 cm, as well as the times required to acquire the 2 image modalities.
RESULTS: The mean and standard deviation of shift discrepancies between MV-topogram and MVCT were 0.74 ± 2.08, -0.09 ± 4.46, and 0.45 ± 3.57 mm in the LAT, longitudinal, and vertical directions, respectively. The MVCT imaging doses measured were 14.74 to 26.92 times higher than the MV-topogram doses, depending on depth. On average, MV-topograms with a mean scan length of 50 cm achieved a 5-fold image acquisition time savings over MVCT, with a mean scan length of 38 cm.
CONCLUSIONS: MV-topograms has the potential to provide alignment performance equivalent to that of MVCT for patients with mesothelioma, with a significant reduction in imaging dose and acquisition time.
METHODS AND MATERIALS: Twelve patients were enrolled in an ongoing institutional review board approved clinical trial at our institute. Patients were set up with a clinical protocol using red lasers. Anteroposterior (AP) and lateral (LAT) MV-topograms were acquired using gantry angles of 0°/90° with a 1 mm collimator opening, all multileaf collimator leaves open, a couch speed of 4 cm/s, and a 12.5-second scanning time. Routine MVCT scans were performed immediately afterward. The MV-topograms were reconstructed and enhanced using contrast-limited adaptive histogram equalization. Anteroposterior and LAT kilovoltage digital reconstructed topogram images were reconstructed based on TomoTherapy geometry from computed tomography simulation scans. Registrations between MV-topograms and kilovoltage-digital reconstructed topogram images were performed manually, and patients' daily shifts were recorded. Results were compared against the corresponding daily MVCT shifts. MV-topogram and MVCT doses were measured and recorded using an ion chamber on a cheese phantom with depths between 1 and 14 cm, as well as the times required to acquire the 2 image modalities.
RESULTS: The mean and standard deviation of shift discrepancies between MV-topogram and MVCT were 0.74 ± 2.08, -0.09 ± 4.46, and 0.45 ± 3.57 mm in the LAT, longitudinal, and vertical directions, respectively. The MVCT imaging doses measured were 14.74 to 26.92 times higher than the MV-topogram doses, depending on depth. On average, MV-topograms with a mean scan length of 50 cm achieved a 5-fold image acquisition time savings over MVCT, with a mean scan length of 38 cm.
CONCLUSIONS: MV-topograms has the potential to provide alignment performance equivalent to that of MVCT for patients with mesothelioma, with a significant reduction in imaging dose and acquisition time.