Choi Gwang Jin, Gwon Tae Mok, Kim Doo Hee, Park Junbeom, Kim Seung Min, Oh Seung Ha, Lim Yoonseob, Jun Sang Beom, Kim Sung June
Аннотация
OBJECTIVE: It is known that the insertion of the intracochlear electrode is critical procedure because the damage around cochlear structures can deteriorate hearing restoration. To reduce the trauma during the electrode insertion surgery, we developed a thin and flexible intracochlear electrode array constructed with carbon nanotube (CNT) bundles.
METHODS: Each CNT bundle was used for an individual electrode channel after coated with parylene C for insulation. By encapsulating eight CNT bundles with silicone elastomer, an 8-channel intracochlear electrode array was fabricated. The mechanical and electrochemical characteristics were assessed to evaluate the flexibility and feasibility of the electrode as a stimulation electrode. The functionality of the electrode was confirmed by electrically evoked auditory brainstem responses (eABR) recorded from a rat.
RESULTS: The proposed electrode has a thickness of 135 μm at the apex and 395 μm at the base. It was demonstrated that the CNT bundle-based electrodes require 6-fold the lower insertion force than metal wire-based electrodes. The electrode impedance and the cathodic charge storage capacitance (CSCc) were 2.70 kΩ ∠-20.4° at 1 kHz and - 708 mC/cm2, respectively. The eABR waves III and V were observed when stimulation current is greater than 50 μA.
CONCLUSION: A thin and flexible CNT bundle-based intracochlear electrode array was successfully developed. The feasibility of the proposed electrode was shown in terms of mechanical and electrochemical characteristics. A proposed CNT bundle-based intracochlear electrode may reduce the risk of trauma during electrode insertion surgery.
METHODS: Each CNT bundle was used for an individual electrode channel after coated with parylene C for insulation. By encapsulating eight CNT bundles with silicone elastomer, an 8-channel intracochlear electrode array was fabricated. The mechanical and electrochemical characteristics were assessed to evaluate the flexibility and feasibility of the electrode as a stimulation electrode. The functionality of the electrode was confirmed by electrically evoked auditory brainstem responses (eABR) recorded from a rat.
RESULTS: The proposed electrode has a thickness of 135 μm at the apex and 395 μm at the base. It was demonstrated that the CNT bundle-based electrodes require 6-fold the lower insertion force than metal wire-based electrodes. The electrode impedance and the cathodic charge storage capacitance (CSCc) were 2.70 kΩ ∠-20.4° at 1 kHz and - 708 mC/cm2, respectively. The eABR waves III and V were observed when stimulation current is greater than 50 μA.
CONCLUSION: A thin and flexible CNT bundle-based intracochlear electrode array was successfully developed. The feasibility of the proposed electrode was shown in terms of mechanical and electrochemical characteristics. A proposed CNT bundle-based intracochlear electrode may reduce the risk of trauma during electrode insertion surgery.