Axillary Temperature, as Recorded by the iThermonitor WT701, Well Represents Core Temperature in Adults Having Noncardiac Surgery.

Pei Lijian, Huang Yuguang, Mao Guangmei, Sessler Daniel I

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Core temperature can be accurately measured from the esophagus or nasopharynx during general anesthesia, but neither site is suitable for neuraxial anesthesia. We therefore determined the precision and accuracy of a novel wireless axillary thermometer, the iThermonitor, to determine its suitability for use during neuraxial anesthesia and in other patients who are not intubated.

METHODS

We enrolled 80 adults having upper abdominal surgery with endotracheal intubation. Intraoperative core temperature was measured in distal esophagus and was estimated at the axilla with a wireless iThermonitor WT701 (Raiing Medical, Boston MA) at 5-minute intervals. Pairs of axillary and reference distal esophageal temperatures were compared and summarized using linear regression and repeated-measured Bland-Altman methods. We a priori determined that the iThermonitor would have clinically acceptable accuracy if most estimates were within ±0.5°C of the esophageal reference, and suitable precision if the limits of agreement were within ±0.5°C.

RESULTS

There were 3339 sets of paired temperatures. Axillary and esophageal temperatures were similar, with a mean difference (esophageal minus axillary) of only 0.14°C ± 0.26°C (standard deviation). The Bland-Altman 95% limits of agreement were reasonably narrow, with the estimated upper limit at 0.66°C and the lower limit at -0.38°C, thus ±0.52°C, indicating good agreement across the range of mean temperatures from 34.9°C to 38.1°C. The absolute difference was within 0.5°C in 91% of the measurements (95% confidence interval, 88%-93%).

CONCLUSIONS

Axillary temperature, as recorded by the iThermonitor WT701, well represents core temperature in adults having noncardiac surgery and thus appears suitable for clinical use.

Для цитирования: Pei Lijian, Huang Yuguang, Mao Guangmei, Sessler Daniel I. Axillary Temperature, as Recorded by the iThermonitor WT701, Well Represents Core Temperature in Adults Having Noncardiac Surgery.. Anesthesia and analgesia. 2018. DOI: 10.1213/ane.0000000000002706.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.