Assessment of chest CT at CTDIvol less than 1 mGy with iterative reconstruction techniques.

Padole Atul, Digumarthy Subba, Flores Efren, Madan Rachna, Mishra Shelly, Sharma Amita, Kalra Mannudeep K

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To assess the image quality of chest CT reconstructed with image-based iterative reconstruction (SafeCT; MedicVision®, Tirat Carmel, Israel), adaptive statistical iterative reconstruction (ASIR; GE Healthcare, Waukesha, WI) and model-based iterative reconstruction (MBIR; GE Healthcare, Waukesha, WI) techniques at CT dose index volume (CTDIvol) <1 mGy.

METHODS

In an institutional review board-approved study, 25 patients gave written informed consent for acquisition of three reduced dose (0.25-, 0.4- and 0.8-mGy) chest CT after standard of care CT (8 mGy) on a 64-channel multidetector CT (MDCT) and reconstructed with SafeCT, ASIR and MBIR. Two board-certified thoracic radiologists evaluated images from the lowest to the highest dose of the reduced dose CT series and subsequently for standard of care CT.

RESULTS

Out of the 182 detected lesions, the missed lesions were 35 at 0.25, 24 at 0.4 and 9 at 0.8 mGy with SafeCT, ASIR and MBIR, respectively. The most missed lesions were non-calcified lung nodules (NCLNs) 25/112 (<5 mm) at 0.25, 18/112 (<5 mm) at 0.4 and 3/112 (<4 mm) at 0.8 mGy. There were 78%, 84% and 97% lung nodules detected at 0.25, 0.4 and 0.8 mGy, respectively regardless of iterative reconstruction techniques (IRTs), Most mediastinum structures were not sufficiently seen at 0.25-0.8 mGy.

CONCLUSION

NCLNs can be missed in chest CT at CTDIvol of <1 mGy (0.25, 0.4 and 0.8 mGy) regardless of IRTs. The most lung nodules (97%) were detected at CTDIvol of 0.8 mGy. The most mediastinum structures were not sufficiently seen at 0.25-0.8 mGy. Advances in knowledge: NCLNs can be missed regardless of IRTs in chest CT at CTDIvol of <1 mGy. The performance of ASIR, SafeCT and MBIR was similar for lung nodule detection at 0.25, 0.4 and 0.8 mGy.

Для цитирования: Padole Atul, Digumarthy Subba, Flores Efren, Madan Rachna, Mishra Shelly, Sharma Amita, Kalra Mannudeep K. Assessment of chest CT at CTDIvol less than 1 mGy with iterative reconstruction techniques.. The British journal of radiology. 2017. DOI: 10.1259/bjr.20160625.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.