Preventing Facial Nerve Stimulation by Triphasic Pulse Stimulation in Cochlear Implant Users: Intraoperative Recordings.

Bahmer Andreas, Adel Youssef, Baumann Uwe

СтатьяПеревод пока не подготовлен

Аннотация

HYPOTHESIS: Triphasic pulse stimulation of the auditory nerve can prevent unintended facial nerve stimulation (FNS) due to a different electromyographic (EMG) input-output function compared with biphasic pulses.

BACKGROUND

FNS is sometimes observed in cochlear implant (CI) users as an unpleasant side effect of electrical stimulation using biphasic pulse patterns (BPP). Clinical remedies to alleviate FNS are 1) to extend stimulus phase duration or 2) to completely deactivate the electrode. In some cases, these options do not provide sufficient FNS reduction or are detrimental to subject performance. Stimulation using triphasic pulse patterns (TPP) has been shown to prevent FNS more effectively, yet the underlying mechanism remains unclear.

METHODS

EMG potentials of muscles innervated by the facial nerve (orbicularis oculi and oris muscles) were recorded to quantitatively compare the effect of BPP and TPP stimulation on FNS. Recordings were conducted in five subjects during CI surgery. In two exemplary cases, different leading phase polarities in alternating and non-alternating order were tested.

RESULTS

Compared with our previous study in awake patients using surface electrodes (Bahmer and Baumann, 2016), intraoperative recordings using subdermal electrodes showed lower noise content and allowed higher sampling resolution. While inter-subject variation remained high, intra-subject results for different electrode positions were comparable: FNS was strongly reduced for cathodic-first TPP stimulation. In contrast, exemplary cases showed little reduction for anodic-first TPP as well as for alternating stimulation.

CONCLUSION

FNS in CI users can be reduced using TPP stimulation, but the ameliorative effect appears to be dependent on the leading stimulus polarity.

Для цитирования: Bahmer Andreas, Adel Youssef, Baumann Uwe. Preventing Facial Nerve Stimulation by Triphasic Pulse Stimulation in Cochlear Implant Users: Intraoperative Recordings.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2017. DOI: 10.1097/mao.0000000000001603.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.