Clinical Evaluation of a Combined Deep Learning-Reconstructed Readout-Segmented Echo-Planar Imaging and Water-Excitation Spectral Fat-Saturation Protocol for Breast Diffusion-Weighted Imaging at 3T Breast MRI.

Choi Jung Min, Lim Soyeoun, Choi Eun Jung, Paek MunYoung, Liu Wei, Bang Minseo, Byon Jung Hee

СтатьяПеревод пока не подготовлен

Аннотация

Objectives

This study evaluates the protocol-level image quality and quantitative diffusion metrics of a clinically implemented deep-learning-reconstructed readout-segmented echo-planar imaging protocol with water-excitation spectral fat saturation (DL-rs-EPI with WEXfs) compared with conventional rs-EPI using spectral attenuated inversion recovery (SPAIR) at 3 T.

Methods

Overall, 80 patients underwent breast magnetic resonance imaging (MRI) with both conventional rs-EPI with SPAIR and DL-rs-EPI with WEXfs protocols (b-values: 0, 800, and 1200 s/mm2). ROI-based relative image-quality metrics, including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and lesion contrast, were assessed at b = 800 and b = 1200 s/mm2; apparent diffusion coefficient (ADC) values were calculated using multi-b-value data. Fat suppression, background diffusion signal, lesion conspicuity, and artifact severity were qualitatively evaluated. A temperature-controlled diffusion phantom (CaliberMRI) was scanned; ADC values were compared with reference values at 24 °C.

Results

DL-rs-EPI with WEXfs demonstrated higher ROI-based relative SNR estimates (b800: 5.79 vs. 5.28; b1200: 5.41 vs. 4.94; p < 0.001) and CNR estimates (b800: 3.35 vs. 3.12, p = 0.024; b1200: 3.67 vs. 3.37, p = 0.001), with unchanged lesion contrast. Tumor ADC values were comparable between protocols, whereas normal fibroglandular tissue ADC values were slightly higher, and ADC contrast increased with DL-rs-EPI with WEXfs. Phantom ADC values from both protocols closely matched reference values at 24 °C, without significant differences. DL-rs-EPI with WEXfs demonstrated more homogeneous fat suppression and reduced background diffusion signal, with comparable lesion conspicuity and artifact severity.

Conclusions

The combined DL-rs-EPI with WEXfs protocol demonstrated improved qualitative and relative quantitative image quality while preserving tumor ADC measurements. As a protocol-level evaluation, these composite improvements support its clinical feasibility for high-quality breast DWI without implying the isolated effect of DL reconstruction alone.

Для цитирования: Choi Jung Min, Lim Soyeoun, Choi Eun Jung, Paek MunYoung, Liu Wei, Bang Minseo, Byon Jung Hee. Clinical Evaluation of a Combined Deep Learning-Reconstructed Readout-Segmented Echo-Planar Imaging and Water-Excitation Spectral Fat-Saturation Protocol for Breast Diffusion-Weighted Imaging at 3T Breast MRI.. Diagnostics (Basel, Switzerland). 2026. DOI: 10.3390/diagnostics16131958.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.