BMC oral health · 2023
Quantitative and qualitative assessment of the wear pattern of two attachment systems of dissimilar materials for mandibular implant-retained overdentures: an in-vitro study.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ramadan Rania E, Mohamed Faten S, Abd-Ellah Mervat E
Аннотация
BACKGROUND: Attachment material is one of the contributing factors to the degree of wear of the attachment components in mandibular implant-retained overdentures. The purpose of this in vitro study was to compare the wear behavior of 2 different attachment systems of dissimilar materials in mandibular implant-retained overdentures by qualitative and quantitative methods.
METHODS: Two attachment systems of different materials were utilized (n = 16); Titach (Dental Evolutions Inc, Beverly Hills, CA, USA) with a titanium-to-titanium interface and Locator R-Tx (Zest Anchors Inc, Escondido, CA, USA) with a titanium-to-nylon interface. One thousand cycles of overdenture insertion and removal simulating 1-year clinical use were performed. All matrices were removed from the overdentures and all patrices were unscrewed from the implants for wear assessment quantitively using a stereomicroscope and qualitatively using a scanning electron microscope. Data were analyzed by using an independent sample t test.
RESULTS: After cyclic loading, stereomicroscopic findings showed that the Titach group had statistically significant higher wear value than the Locator R-Tx group (p < 0.001). Moreover, scanning electron microscopy images showed noticeable abrasion in Titach patrix at the area of highest convexity. However, the Locator R-Tx matrix displayed an apparent tear of rubber inserts.
CONCLUSIONS: Titach attachment with the titanium-to-titanium interface revealed more wear than Locator R-Tx attachment with the titanium-to-nylon interface. Thus, the type of attachment material influences the degree of wear of the attachment components.
METHODS: Two attachment systems of different materials were utilized (n = 16); Titach (Dental Evolutions Inc, Beverly Hills, CA, USA) with a titanium-to-titanium interface and Locator R-Tx (Zest Anchors Inc, Escondido, CA, USA) with a titanium-to-nylon interface. One thousand cycles of overdenture insertion and removal simulating 1-year clinical use were performed. All matrices were removed from the overdentures and all patrices were unscrewed from the implants for wear assessment quantitively using a stereomicroscope and qualitatively using a scanning electron microscope. Data were analyzed by using an independent sample t test.
RESULTS: After cyclic loading, stereomicroscopic findings showed that the Titach group had statistically significant higher wear value than the Locator R-Tx group (p < 0.001). Moreover, scanning electron microscopy images showed noticeable abrasion in Titach patrix at the area of highest convexity. However, the Locator R-Tx matrix displayed an apparent tear of rubber inserts.
CONCLUSIONS: Titach attachment with the titanium-to-titanium interface revealed more wear than Locator R-Tx attachment with the titanium-to-nylon interface. Thus, the type of attachment material influences the degree of wear of the attachment components.