m+Medelevator.RU
← Поиск публикаций
Clinical and experimental dental research · 2023

Two- and three-piece implants to boost data generation in preclinical in vivo research-A short technical report.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Stavropoulos Andreas, Bellon Benjamin, Pipenger Benjamin, Andersen Ole Z

Аннотация

The purpose of this technical report is to present two novel experimental implant designs to boost data generation in preclinical in vivo research. Specifically, the report describes the rationale and the components of (1) a two-piece experimental implant suitable for a small animal platform (e.g., the rabbit femur/tibial epiphysis model), consisting of a threaded apical- and a coronal cylindrical piece, which is intended for collecting two types of biomechanical data, and (2) a three-piece experimental implant suitable for a large animal platform (e.g., the mini-pig mandible model), consisting of an apical "wound chamber", which allows the collection of histological/histomorphometrical data, and a middle threaded and coronal cylindrical piece, which also allow the collection of two types of biomechanical data. The increased volume of information generated from a single experiment in a small animal platform, using the proposed two-piece implant design, may assist in a more qualified decision-making process, on whether it is relevant to proceed to further assessment using a large animal platform. Furthermore, the increased volume of information generated in a single animal experiment either in a small or large animal platform, using the proposed two- and three-piece implants, respectively, likely decreases the number of animals otherwise needed for collecting the same information with standard one-piece implants and, thus, contributes to the reduction/refinement elements of the 3R principle.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты