Confirmation of Essure microinsert tubal coil placement with conventional and volume-contrast imaging three-dimensional ultrasound.

Thiel John A, Suchet Ian B, Lortie Karine

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine the accuracy of ultrasound in the assessment of proximal fallopian tube positioning of the Essure microinsert coil 3 months after postprocedure.

DESIGN

Prospective cohort study (Canadian Task Force classification II-2).

SETTING

Reproductive-age women in a tertiary care hospital.

PATIENT(S): Reproductive-age women presenting with a request for permanent contraception.

INTERVENTION(S): Hysteroscopic sterilization with the Essure microinsert coil and conventional or volume-contrast three-dimensional (3D) ultrasound imaging 3 months after the procedure.

MAIN OUTCOME MEASURE(S): Coil position on ultrasound.

RESULT(S): Forty-eight of the 50 patients had successful placement of the Essure coils, and three patients required a second attempt on one tube. Conventional or volume-contrast (3D) ultrasound showed proper positioning of the coils within the proximal fallopian tube in 42 women (84%); five women (10%) required hysterosalpingogram to show appropriate positioning. Two patients (4%) required laparoscopic tubal sterilization, and one patient (2%) was lost to follow-up.

CONCLUSION(S): Transvaginal ultrasound is an acceptable method of confirming proper placement of the Essure microinsert coil within the proximal fallopian tube 3 months after the procedure.

Для цитирования: Thiel John A, Suchet Ian B, Lortie Karine. Confirmation of Essure microinsert tubal coil placement with conventional and volume-contrast imaging three-dimensional ultrasound.. Fertility and sterility. 2005. DOI: 10.1016/j.fertnstert.2005.01.135.