Results of a multicenter 4D computed tomography quality assurance audit: Evaluating image accuracy and consistency.

Burghelea Manuela, Bakkali Tahiri Jinane, Dhont Jennifer, Kyndt Martin, Gulyban Akos, Szkitsak Juliane, Bogaert Evelien, van Gestel Dirk, Reynaert Nick

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: 4D Computed Tomography (4DCT) technology captures the location and movement of tumors and nearby organs at risk over time. In this study, a multi-institutional multi-vendor 4DCT audit was initiated to assess the accuracy of current imaging protocols.

MATERIALS AND

METHODS

Twelve centers, including thirteen scanners performed a 4DCT acquisition of a dynamic thorax phantom using the institution's own protocol with the in-house breathing monitoring system. Five regular and three irregular breathing patterns were used. Image acquisition and reconstruction were followed by automated image analysis with our in-house developed 4DCT QA program (QAMotion). CT number accuracy, volume deviation, amplitude deviation, and spatial integrity were assessed per pattern using both the segmented volumes and line profiles.

RESULTS

Regular breathing curves showed relatively accurate results across all institutions, with mean volume and CT number deviations and median amplitude deviation below 2%, 5 HU and 2 mm, respectively. Results obtained for irregular patterns showed more variation across the institutions. Volume and CT number deviations co-occurred with a blurring of the sphere, interpolation, or double-structure artifacts that were confirmed through the line profiles. For some of the irregular patterns, amplitude deviations up to 6 mm were observed. Maximum Intensity Projection (MaxIP) correctly captured the applied motion amplitude with deviations across all institutions within 2 mm except for double amplitude pattern.

CONCLUSIONS

All centers invited to participate in the audit responded positively, highlighting the need for a comprehensive yet easy-to-execute 4DCT quality assurance program. The largest variances between the results from one institution to another confirmed that a standardized 4DCT audit is warranted.

Для цитирования: Burghelea Manuela, Bakkali Tahiri Jinane, Dhont Jennifer, Kyndt Martin, Gulyban Akos, Szkitsak Juliane, Bogaert Evelien, van Gestel Dirk, Reynaert Nick. Results of a multicenter 4D computed tomography quality assurance audit: Evaluating image accuracy and consistency.. Physics and imaging in radiation oncology. 2023. DOI: 10.1016/j.phro.2023.100479.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.