A Method to Improve the Accuracy for the Comparison of Consecutively Determined Upper Limb Cross-Sectional Area Profiles of Test Subjects: Impact on (Segmental) Limb Volume and Local Limb Circumference.
Аннотация
Background
Chronic edema management depends on measuring small volume changes over time. Therefore, a highly sensitive, accurate, and reliable technique is needed to objectively judge sequential measurements. Methods and
Results
The prototype of the Peracutus Aqua Meth (PAM), a new volumetric measuring device, was used in an experimental study. Thirty-nine healthy test subjects were measured three times. The measuring principle of the PAM is based on obtaining a profile of cross-sectional areas along the length of an object or limb. Besides total arm volumes, the volume of any chosen segment can be determined. The utmost tip of the middle finger appeared to be an unreliable reference point. Instead, the concept of "hand median" was introduced to align and compare profiles of cross-sectional areas of sequential measurements. Using the hand median strongly improved the measuring accuracy, rendering standard deviation values of 0.4%-1.3% for (segmental) volume determination.
Conclusions
Measurements with the PAM show that accurate local/segmental volume determination of a limb is possible. Standard deviations of less than 1.3% are easily obtained for cross-sectional area and volume determination.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.