DOI 10.1097/rct.0000000000001538

Медицинская библиотека · Journal of computer assisted tomography · 2023

Usefulness of Ag Additional Filter on Image Quality and Radiation Dose for Low-Dose Chest Computed Tomography.

Kawamoto Keishin, Sato Hideyuki, Kogure Yosuke

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

This study aimed to evaluate the effect of a silver (Ag) additional filter on dose characteristics and image quality in low-dose chest computed tomography (CT).

METHODS

A dose evaluation phantom, physical evaluation phantom, and chest phantom were scanned with and without an Ag additional filter. The doses were adjusted so that the displayed the volume CT dose indexes (CTDI vol ) were from 0.3 to 1.6 mGy. For dose characteristics, the spectrum of photon energies and the measured CTDI vol were calculated for each scanning condition. For task-based image quality analysis, task transfer function, noise power spectrum, and system performance were evaluated. Streak artifacts, image noise, and contrast-to-noise ratio were quantified using a chest phantom.

RESULTS

With the Ag additional filter, mean energy was 22% higher and the CTDI vol was approximately 30% lower than those without the Ag additional filter. The task transfer function and noise power spectrum with the Ag additional filter were lower than those without the Ag additional filter. The system performance with the Ag additional filter was similar to that without the Ag additional filter. The Ag additional filter reduced streak artifact near the lung apex and image noise in the lung fields. The contrast-to-noise ratio was slightly higher with the Ag additional filter than that without the Ag additional filter.

CONCLUSIONS

The output dose and spatial resolution with the Ag additional filter were lower than those without the Ag additional filter. However, this filter helped reduce the radiation dose, image noise, and streak artifacts, particularly when scanning at ultralow doses.

Для цитирования: Kawamoto Keishin, Sato Hideyuki, Kogure Yosuke. Usefulness of Ag Additional Filter on Image Quality and Radiation Dose for Low-Dose Chest Computed Tomography.. Journal of computer assisted tomography. 2023. DOI: 10.1097/rct.0000000000001538.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.