Cancer cell international · 2023
Development and validation of sensitive BCR::ABL1 fusion gene quantitation using next-generation sequencing.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lee Hyeonah, Seo Jieun, Shin Saeam, Lee Seung-Tae, Choi Jong Rak
Аннотация
BACKGROUND: BCR::ABL1 fusion has significant prognostic value and is screened for chronic myeloid leukemia (CML) disease monitoring as a part of routine molecular testing. To overcome the limitations of the current standard real-time quantitative polymerase chain reaction (RQ-PCR), we designed and validated a next-generation sequencing (NGS)-based assay to quantify BCR::ABL1 and ABL1 transcript copy numbers.
METHODS: After PCR amplification of the target sequence, deep sequencing was performed using an Illumina Nextseq 550Dx sequencer and in-house-designed bioinformatics pipeline. The Next-generation Quantitative sequencing (NQ-seq) assay was validated for its analytical performance, including precision, linearity, and limit of detection, using serially diluted control materials. A comparison with conventional RQ-PCR was performed with 145 clinical samples from 77 patients.
RESULTS: The limit of detection of the NQ-seq was the molecular response (MR) 5.6 [BCR::ABL1 0.00028% international scale (IS)]. The NQ-seq exhibited excellent precision and linear range from MR 2.0 to 5.0. The IS value from the NQ-seq was highly correlated with conventional RQ-PCR.
CONCLUSIONS: We conclude that the NQ-seq is an effective tool for monitoring BCR::ABL1 transcripts in CML patients with high sensitivity and reliability. Prospective assessment of the unselected large series is required to validate the clinical impact of this NGS-based monitoring strategy.
METHODS: After PCR amplification of the target sequence, deep sequencing was performed using an Illumina Nextseq 550Dx sequencer and in-house-designed bioinformatics pipeline. The Next-generation Quantitative sequencing (NQ-seq) assay was validated for its analytical performance, including precision, linearity, and limit of detection, using serially diluted control materials. A comparison with conventional RQ-PCR was performed with 145 clinical samples from 77 patients.
RESULTS: The limit of detection of the NQ-seq was the molecular response (MR) 5.6 [BCR::ABL1 0.00028% international scale (IS)]. The NQ-seq exhibited excellent precision and linear range from MR 2.0 to 5.0. The IS value from the NQ-seq was highly correlated with conventional RQ-PCR.
CONCLUSIONS: We conclude that the NQ-seq is an effective tool for monitoring BCR::ABL1 transcripts in CML patients with high sensitivity and reliability. Prospective assessment of the unselected large series is required to validate the clinical impact of this NGS-based monitoring strategy.