The impact of margin reduction on radiation dose distribution of ultra-hypofractionated prostate radiotherapy utilizing a 1.5-T MR-Linac.

Onal Cem, Efe Esma, Bozca Recep, Yavas Cagdas, Yavas Guler, Arslan Gungor

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

We examined the effects of reducing the planning target volume (PTV) margin in MR-guided radiotherapy (MRgRT) on the distribution of radiation dose to target volumes and organs-at-risk (OARs). Thus, we compared MR-Linac (MRL) plans with and without reduced margin and intensity-modulated radiotherapy (IMRT) plan with conventional linac for low-risk prostate cancer patients receiving 36.25 Gy in five fractions of ultra-hypofractionated radiation therapy.

MATERIALS AND

METHODS

Twenty low-risk prostate cancer patients treated with 1.5 T MR-Linac were evaluated. The same planning CT images were used for four plans: the MRL-R plan with reduced margin planning target volume (PTV-R) and the MRL-N plan with normal margin PTV (PTV-N), which is also used for IMRT plan. In four plans, PTV doses, organs-at-risk (OARs) doses, the homogeneity index (HI), and monitor units were compared.

RESULTS

All plans met the criteria for PTV coverage and OARs dose constraints. The maximum and mean PTV doses were significantly higher in the MRL-R and MRL-N plans compared to the IMRT plan. The HI was lowest in the IMRT plan (0.040 ± 0.013) and highest in the MRL-N plan (0.055 ± 0.012; p < 0.001). There was no significant difference in the PTV dosimetric parameters between the MRL-R and the MRL-N plans. The high doses in the rectum was significantly lower in the MRL-R compared to other plans. The bladder V36.25 Gy was significantly lower in the MRL-R plan (2.43 ± 1.87 Gy) compared to MRL-N (4.50 ± 2.42 Gy; p < 0.001), and IMRT plans (4.76 ± 2.77 Gy; p < 0.001). There was no significant difference in the low-dose volumes of the body, maximum femur doses, or monitor units across each plan.

CONCLUSIONS

Ultra-hypofractionated MR-guided RT with 1.5 T MRL is dosimetrically feasible for patients with prostate cancer. The improved soft tissue contrast and the online adaptive plan for 1.5 T MR-Linac allows for PTV margin reduction resulted in a significant dose reduction in OARs.

Для цитирования: Onal Cem, Efe Esma, Bozca Recep, Yavas Cagdas, Yavas Guler, Arslan Gungor. The impact of margin reduction on radiation dose distribution of ultra-hypofractionated prostate radiotherapy utilizing a 1.5-T MR-Linac.. Journal of applied clinical medical physics. 2023. DOI: 10.1002/acm2.14179.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.