Chest CT in patients after lung transplantation: A retrospective analysis to evaluate impact on image quality and radiation dose using spectral filtration tin-filtered imaging.

Wressnegger Alexander, Prosch Helmut, Moser Bernhard, Klepetko Walter, Jaksch Peter, Lambers Christopher, Hoetzenecker Konrad, Schestak Christian, De Bettignies Albert, Beer Lucian, Apfaltrer Georg, Ringl Helmut, Apfaltrer Paul

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The purpose of this study was to investigate the impact of a 150kV spectral filtration chest imaging protocol (Sn150kVp) combined with advanced modeled iterative reconstruction (ADMIRE) on radiation dose and image quality in patients after lung-transplantation.

METHODS

This study included 102 patients who had unenhanced chest-CT examinations available on both, a second-generation dual-source CT (DSCT) using standard protocol (100kVp, filtered-back-projection) and, on a third-generation DSCT using Sn150kVp protocol with ADMIRE. Signal-to-noise-ratio (SNR) was measured in 6 standardized regions. A 5-point Likert scale was used to evaluate subjective image quality. Radiation metrics were compared.

RESULTS

The mean time interval between the two acquisitions was 1.1±0.7 years. Mean-volume-CT-dose-index, dose-length-product and effective dose were significantly lower for Sn150kVp protocol (2.1±0.5mGy;72.6±16.9mGy*cm;1.3±0.3mSv) compared to 100kVp protocol (6.2±1.8mGy;203.6±55.6mGy*cm;3.7±1.0mSv) (p<0.001), equaling a 65% dose reduction. All studies were considered of diagnostic quality. SNR measured in lung tissue, air inside trachea, vertebral body and air outside the body was significantly higher in 100kVp protocol compared to Sn150kVp protocol (12.5±2.7vs.9.6±1.5;17.4±3.6vs.11.8±1.8;0.7±0.3vs.0.4±0.2;25.2±6.9vs.14.9±3.3;p<0.001). SNR measured in muscle tissue was significantly higher in Sn150kVp protocol (3.2±0.9vs.2.6±1.0;p<0.001). For SNR measured in descending aorta there was a trend towards higher values for Sn150kVp protocol (2.8±0.6 vs. 2.7±0.9;p = 0.3). Overall SNR was significantly higher in 100kVp protocol (5.0±4.0vs.4.0±4.0;p<0.001). On subjective analysis both protocols achieved a median Likert rating of 1 (25th-75th-percentile:1-1;p = 0.122). Interobserver agreement was good (intraclass correlation coefficient = 0.73).

CONCLUSIONS

Combined use of 150kVp tin-filtered chest CT protocol with ADMIRE allows for significant dose reduction while maintaining highly diagnostic image quality in the follow up after lung transplantation when compared to a standard chest CT protocol using filtered back projection.

Для цитирования: Wressnegger Alexander, Prosch Helmut, Moser Bernhard, Klepetko Walter, Jaksch Peter, Lambers Christopher, Hoetzenecker Konrad, Schestak Christian, De Bettignies Albert, Beer Lucian, Apfaltrer Georg, Ringl Helmut, Apfaltrer Paul. Chest CT in patients after lung transplantation: A retrospective analysis to evaluate impact on image quality and radiation dose using spectral filtration tin-filtered imaging.. PloS one. 2020. DOI: 10.1371/journal.pone.0228376.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.