Intensive care medicine experimental · 2020

In vitro comparison between inspiration synchronized and continuous vibrating mesh nebulizer during trans-nasal aerosol delivery.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Li Jie, Wu Wei, Fink James B

Аннотация

BACKGROUND: Compared to continuous vibrating mesh nebulizer (VMN), inspiration synchronized VMN has shown increased inhaled dose during noninvasive ventilation; however, its use during aerosol delivery via high-flow nasal cannula (HFNC) is still unknown.

METHODS: An adult manikin was connected to a dual-chamber model lung, which was driven by a critical care ventilator to simulate spontaneous breathing. A HFNC system was utilized with temperature at 37 ° C while gas flow at 5, 10, 20, 40, and 60 L/min. Inspiration synchronized and continuous aerosol generation were compared at different positions (at the inlet of humidifier vs close to patient). One milliliter of albuterol (2.5 mg/mL) was used in each run (n = 3). Collection filter was placed at the trachea and was removed after each run. Drug was eluted from the filter and assayed with UV spectrophotometry (276 nm).

RESULTS: When nebulizer was placed close to patient, inhaled dose was higher with inspiration synchronized than continuous aerosol generation at all gas flows (p = 0.05) except at 5 L/min. When placed at the inlet of humidifier, compared to continuous, inspiration synchronized aerosol generated higher inhaled dose with gas flow set below 50% of patient inspiratory flow [23.9 (20.6, 28.3)% vs 18.1 (16.7, 19.6)%, p < 0.001], but lower inhaled dose with gas flow set above 50% of patient inspiratory flow [3.5 (2.2, 9.3)% vs 9.9 (8.2, 16.4)%, p = 0.001]. Regardless of breathing pattern, continuous aerosol delivered greater inhaled dose with nebulizer placed at humidifier than close to patient at all gas flows except at 5 L/min.

CONCLUSION: When the HFNC gas flow was set higher than 50% of patient inspiratory flow, no significant advantage was found in inspiration synchronized over continuous aerosol. However, inspiration synchronized aerosol generated 30% more inhaled dose than continuous with gas flow set below 50% of patient inspiratory flow, regardless of nebulizer placement. Continuous nebulizer needs to be placed at the inlet of humidifier.