DOI 10.1177/19322968251315499

Медицинская библиотека · Journal of diabetes science and technology · 2025

A Performance-Based Adaptation Index for Automated Insulin Delivery Systems.

Diaz C Jenny L, Colmegna Patricio, Pryor Elliot, Breton Marc D

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Automated insulin delivery (AID) algorithms can benefit from tuning of their aggressiveness to meet individual needs, as insulin requirements vary among and within users. We introduce the Performance-Based Adaptation Index (PAI), a tool designed to enable automatic adjustment of an AID system aggressiveness based on continuous glucose monitoring (CGM) metrics.

METHODS

PAI integrates two CGM-based metrics-one for hypoglycemia and another for hyperglycemia exposure-over a previous time window into a single index (αθ). We propose two methods to compute αθ: one based on time in range (TIR, 70-180 mg/dL), and the other on glycemic risk indices. Using αθ, we developed a multiplicative strategy to adjust the AID system's aggressiveness, accounting for situations where αθ cannot be reliably calculated. The feasibility of this method was assessed in-silico using the UVA/Padova Type 1 Diabetes Simulator and our full closed-loop algorithm (UVA-model predictive control (MPC)) across five scenarios: optimal tuning (baseline), conservative and aggressive tunings, and temporary and permanent changes in insulin needs. Glycemic outcomes were evaluated from the simulated glucose traces.

RESULTS

Negligible performance variations were observed in the baseline scenario. For the conservative scenario, adjusting αθ improved TIR (35.1% vs 71.8%) and increased total daily insulin (32.1 U vs 41.2 U). Conversely, for the aggressive scenario, it reduced hypoglycemia exposure (TBR: 2.6% vs 1.4%) and overall insulin usage (45.6 U vs 43.0 U).

CONCLUSION

In-silico results demonstrated the safety and efficacy of using PAI to automatically tune the UVA-MPC controller, achieving TIR values above 70% under fully closed-loop conditions and across various physiological states. Clinical validation of these results is warranted.

Для цитирования: Diaz C Jenny L, Colmegna Patricio, Pryor Elliot, Breton Marc D. A Performance-Based Adaptation Index for Automated Insulin Delivery Systems.. Journal of diabetes science and technology. 2025. DOI: 10.1177/19322968251315499.