Аннотация

PURPOSE: To compare outcomes of percutaneous robot-assisted uterine fibroid embolization (UFE) using two different endovascular robotic catheters.

METHODS

Twenty-one patients with a symptomatic uterine fibroid were prospectively enrolled in a single-center study to be treated with a percutaneous robot-assisted embolization using the Magellan system. Fourteen patients were treated using a first generation steerable robotic catheter, version 1.0 (group 1), and seven were treated using the new version 1.1 (group 2). Demographic, pathologic, and procedural variables were recorded. Dose Area Product (DAP) and physician equivalent doses were registered for each procedure. Procedural related complications and clinical midterm outcomes were also evaluated.

RESULTS

Successful robot-assisted UFE was obtained in eight patients (57.1%) in group 1 and 7 patients (100%) in group 2 (p = 0.01). A successful robot-assisted catheterization of the internal iliac artery anterior branch was performed in all patients of both groups. Median selective target vessel catheterization time was 21.0 ± 12.8 vs 13.4 ± 7 min (p = 0.04) and total fluoroscopy time was 30.3 ± 11.2 vs 19.3 ± 5.9 min, respectively, in group 1 and 2. Mean DAP decreased from 18472.6 ± 15622 to 5469.1 ± 4461.0 cGy·cm2 (p = 0.04). All patients obtained a symptoms relief at 6 months follow-up.

CONCLUSIONS

Robot-assisted uterine fibroid embolization is safe and effective. New version of steerable robotic catheter allows performing a faster procedure without related adverse events compared to old version.

Для цитирования: Del Giudice C, Pellerin O, Nouri Neville M, Amouyal G, Fitton I, Leré-Déan C, Sapoval M. Comparison of Two Endovascular Steerable Robotic Catheters for Percutaneous Robot-Assisted Fibroid Embolization.. Cardiovascular and interventional radiology. 2017. DOI: 10.1007/s00270-017-1790-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.