Stimulus characteristics of a novel air-based multiple stimulus aesthesiometer.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Mungalsingh Melanie A, Thompson Benjamin, Peterson Sean D, Murphy Paul J

Аннотация

PURPOSE: To identify the stimulus airflow characteristics and confirm the consistency of a novel air jet-based aesthesiometer capable of producing and applying multiple stimuli separated either by time and/or by space.

METHODS: A novel aesthesiometer (Dolphin Aesthesiometer) was designed around a micro-blower under software management. Two nozzle attachments assisted in airflow control (flexible tube 1.6 mm diameter; brass tube 0.5 mm diameter). Four studies that tested the characteristics of the airflow and stimulus consistency were completed: (i) airflow pattern/trajectory, (ii) airflow surface dispersion, (iii) force of airflow across a range of stimulus strengths and (iv) thermal effects on the ocular surface.

RESULTS: Stimulus characteristic studies revealed: (i) airflow is coherent within the expected test distance range for the instrument, and spread rate is constant irrespective of stimulus strength; (ii) airflow dispersion occurs upon encountering a surface and dispersion increases with increasing airflow rate; (iii) a consistent and small force (10-4  N) is applied by the airflow and (iv) repeatable thermal effects occur in relation to the airflow, and the mode of stimulation of the Dolphin aesthesiometer is predominantly thermal in nature.

CONCLUSIONS: These studies confirm the repeatability and consistency of the novel instrument. The device is suitable for measuring corneal sensitivity. The availability of additional air jets allows the application of multiple stimuli to facilitate corneal summation investigations.