MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jclinane.2016.11.005

Медицинская библиотека · Journal of clinical anesthesia · 2016

Intraoperative aberrant bispectral index values due to facial nerve monitoring.

Chakrabarti Dhritiman, Surve Rohini Mrugank, Bs Deepti, Masapu Dheeraj

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Bispectral index is an accepted depth of anaesthesia monitor for guiding intraoperative hypnotic agent administration. Frontalis EMG displayed on BIS monitor may increase due to twitching of frontalis muscle. EMG increases are also known to cause artefactual increases in BIS values. We report a case of artefactual increase of EMG and subsequently BIS values, due to electrical artefact from cranial nerve stimulator being used to identify the facial nerve. An explanation of the effect of stimulator signal on BIS EMG and BIS values has been provided.

Для цитирования: Chakrabarti Dhritiman, Surve Rohini Mrugank, Bs Deepti, Masapu Dheeraj. Intraoperative aberrant bispectral index values due to facial nerve monitoring.. Journal of clinical anesthesia. 2016. DOI: 10.1016/j.jclinane.2016.11.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.