T-Connector Modification for Reducing Recurrent Distal Shunt Failure: Report of 2 Cases.
Аннотация
BACKGROUND AND
IMPORTANCE
Cerebrospinal fluid shunt placement is used to treat the various causes of hydrocephalus by redirecting the cerebrospinal fluid to the body, most commonly from the ventricle to the peritoneum. Distal catheter displacement from the peritoneal cavity can occur as a complication, necessitating reoperation.
CLINICAL PRESENTATION: We report 2 such cases in obese patients involving retropulsion of the distal tubing. To address this complication, we implanted a T-connector to the distal catheter construct.
CONCLUSION
This study supports the use of a T-connector catheter construct to decrease and prevent the possibility of distal peritoneal catheter retropulsion in cases of elevated intra-abdominal pressure, both prophylactically and in revisions.