Londeree William, Davis Konrad, Helman Donald, Abadie Jude
Аннотация
INTRODUCTION: Forward deployed military medical units can provide sophisticated medical care with limited resources. Point-of-Care Testing (POCT) may facilitate care and expedite diagnosis. This study assessed the accuracy of results for POCT for non-serum samples (pleural, peritoneal, and cerebrospinal fluid) using iSTAT and Piccolo hand-held devices compared with results obtained using a hospital chemistry analyzer.
METHODS: Pleural, peritoneal, and cerebrospinal fluids obtained during routine care were simultaneously analyzed on a Vitros 5600 automated clinical chemistry hospital analyzer, iSTAT, and Piccolo POCT devices.
RESULTS: POCT results were highly correlated with the Vitros 5600 for pleural fluid LDH, glucose, and triglycerides (TG); for peritoneal fluid bilirubin, TG, glucose, albumin, and protein; and glucose for cerebrospinal fluid.
CONCLUSION: POCT results for non-serum samples from pleural, peritoneal, and cerebrospinal fluid correlate with standard hospital chemistry analysis. The results of this study demonstrate potential for possible new diagnostic roles for POCT in resource-limited environments.
METHODS: Pleural, peritoneal, and cerebrospinal fluids obtained during routine care were simultaneously analyzed on a Vitros 5600 automated clinical chemistry hospital analyzer, iSTAT, and Piccolo POCT devices.
RESULTS: POCT results were highly correlated with the Vitros 5600 for pleural fluid LDH, glucose, and triglycerides (TG); for peritoneal fluid bilirubin, TG, glucose, albumin, and protein; and glucose for cerebrospinal fluid.
CONCLUSION: POCT results for non-serum samples from pleural, peritoneal, and cerebrospinal fluid correlate with standard hospital chemistry analysis. The results of this study demonstrate potential for possible new diagnostic roles for POCT in resource-limited environments.