INTRODUCTION

Forward deployed military medical units can provide sophisticated medical care with limited resources. Point-of-Care Testing (POCT) may facilitate care and expedite diagnosis. This study assessed the accuracy of results for POCT for non-serum samples (pleural, peritoneal, and cerebrospinal fluid) using iSTAT and Piccolo hand-held devices compared with results obtained using a hospital chemistry analyzer.

METHODS

Pleural, peritoneal, and cerebrospinal fluids obtained during routine care were simultaneously analyzed on a Vitros 5600 automated clinical chemistry hospital analyzer, iSTAT, and Piccolo POCT devices.

RESULTS

POCT results were highly correlated with the Vitros 5600 for pleural fluid LDH, glucose, and triglycerides (TG); for peritoneal fluid bilirubin, TG, glucose, albumin, and protein; and glucose for cerebrospinal fluid.

CONCLUSION

POCT results for non-serum samples from pleural, peritoneal, and cerebrospinal fluid correlate with standard hospital chemistry analysis. The results of this study demonstrate potential for possible new diagnostic roles for POCT in resource-limited environments.

Для цитирования: Londeree William, Davis Konrad, Helman Donald, Abadie Jude. Bodily fluid analysis of non-serum samples using point-of-care testing with iSTAT and Piccolo analyzers versus a fixed hospital chemistry analytical platform.. Hawai'i journal of medicine & public health : a journal of Asia Pacific Medicine & Public Health. 2014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.