MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.bjps.2020.05.040

Медицинская библиотека · Journal of plastic, reconstructive & aesthetic surgery : JPRAS · 2020

Near-infrared spectroscopy for monitoring free jejunal flap.

Yano Akiko, Orihashi Kazumasa, Yoshida Yukitaka, Kuriyama Motone

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Although the free jejunal graft is commonly used for reconstruction after resection of a tumor of the pharynx or cervical esophagus, adequate monitoring for detecting graft failure is not available. We employed near-infrared spectroscopy to measure regional oxygen saturation (rSO2) in the graft.

METHODS

In 25 consecutive cases who underwent reconstructive surgery using a free jejunal graft, the feasibility of postoperative rSO2 monitoring was examined along with the changes in rSO2 values following vascular clamping or reperfusion.

RESULTS

No operative mortality occurred, and except for one case of subcutaneous hematoma that necessitated evacuation surgery, no complications related to surgery or graft failure occurred. Postoperative rSO2 monitoring was feasible for >50 hours in most cases. It mostly remained >55% with a stable hemoglobin index (HbI) which reflects tissue hemoglobin density. A marked increase in the HbI was noted in a patient with hematoma. Intraoperatively, the rSO2 of intact jejunal tissue was >60% in every case but dropped within a few minutes after arterial clamping because of decreased oxygenated hemoglobin concentration. With venous clamping, the HbI was elevated while the rSO2 remained unchanged or was slightly decreased. Upon graft reperfusion, the rSO2 rapidly recovered in all 18 cases because of the recovery of oxygenated hemoglobin concentrations.

CONCLUSIONS

The near-infrared spectroscopic assessment sensitively and accurately reflected the condition of the jejunal graft. It appears to be a promising postoperative method for monitoring graft perfusion. An rSO2 value of 55% appears to be an adequate criterion for ischemia.

Для цитирования: Yano Akiko, Orihashi Kazumasa, Yoshida Yukitaka, Kuriyama Motone. Near-infrared spectroscopy for monitoring free jejunal flap.. Journal of plastic, reconstructive & aesthetic surgery : JPRAS. 2020. DOI: 10.1016/j.bjps.2020.05.040.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.