MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000001867

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2018

Automated Metrics in a Virtual-Reality Myringotomy Simulator: Development and Construct Validity.

Huang Caiwen, Cheng Horace, Bureau Yves, Ladak Hanif M, Agrawal Sumit K

СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The objectives of this study were: 1) to develop and implement a set of automated performance metrics into the Western myringotomy simulator, and 2) to establish construct validity.

STUDY

DESIGN

Prospective simulator-based assessment study.

SETTING

The Auditory Biophysics Laboratory at Western University, London, Ontario, Canada.

PARTICIPANTS

Eleven participants were recruited from the Department of Otolaryngology-Head & Neck Surgery at Western University: four senior otolaryngology consultants and seven junior otolaryngology residents.

INTERVENTIONS: Educational simulation.

MAIN OUTCOME MEASURE: Discrimination between expert and novice participants on five primary automated performance metrics: 1) time to completion, 2) surgical errors, 3) incision angle, 4) incision length, and 5) the magnification of the microscope.

METHODS

Automated performance metrics were developed, programmed, and implemented into the simulator. Participants were given a standardized simulator orientation and instructions on myringotomy and tube placement. Each participant then performed 10 procedures and automated metrics were collected. The metrics were analyzed using the Mann-Whitney U test with Bonferroni correction.

RESULTS

All metrics discriminated senior otolaryngologists from junior residents with a significance of p < 0.002. Junior residents had 2.8 times more errors compared with the senior otolaryngologists. Senior otolaryngologists took significantly less time to completion compared with junior residents. The senior group also had significantly longer incision lengths, more accurate incision angles, and lower magnification keeping both the umbo and annulus in view.

CONCLUSIONS

Automated quantitative performance metrics were successfully developed and implemented, and construct validity was established by discriminating between expert and novice participants.

Для цитирования: Huang Caiwen, Cheng Horace, Bureau Yves, Ladak Hanif M, Agrawal Sumit K. Automated Metrics in a Virtual-Reality Myringotomy Simulator: Development and Construct Validity.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2018. DOI: 10.1097/mao.0000000000001867.