Медицинская библиотека · The International journal of oral & maxillofacial implants · 2023

Histologic Osseointegration Level Comparing Titanium and Zirconia Dental Implants: Meta-analysis of Preclinical Studies.

Remísio Maria João da Silva, Borges Tiago, Castro Filipe, Gehrke Sergio Alexandre, Fernandes Juliana Campos Hasse, Fernandes Gustavo Vicentis de Oliveira

Систематический обзорМетаанализСтатьяПеревод пока не подготовлен

Аннотация

Purpose: To assess the literature comparing histologic levels of osseointegration for titanium vs zirconia dental implants. Materials and

Methods

This systematic review was conducted following the PRISMA guidelines and was registered in PROSPERO (CRD42021236781). Electronic and manual searches were carried out through the PubMed/MEDLINE, PubMed Central, and Embase databases with a platform-specific search strategy combining controlled terms (MeSH and Emtree) and text words. The articles were selected by two independent investigators who evaluated the articles based on the criteria for eligibility.

Results

A total of 17 articles were included. All were preclinical studies. The populations included dogs (27.55%), minipigs (14.28%), rats (14.28%), and rabbits (43.89%); and the implantation site varied among the mandible (36.82%), maxilla (9.04%), tibia (17.64%), skull (10.70%), and femur (25.80%). A total of 370 titanium (Ti) implants and 537 zirconia (Zr) implants were evaluated. The average osseointegration (% bone-to-implant contact) for Zr was 55.51% (17.6% to 89.09%), and for Ti was 58.50% (23.2% to 87.85%). There was no statistical difference between studies at the 2-month follow-up (P = .672), but this difference was significant at 1 and 3 months (P < .001).

Conclusions

Within the limitations of this review, Zr implants had a similar level of osseointegration compared to Ti implants. Nonetheless, because these findings are based on preclinical research, all data must be carefully examined.

Для цитирования: Remísio Maria João da Silva, Borges Tiago, Castro Filipe, Gehrke Sergio Alexandre, Fernandes Juliana Campos Hasse, Fernandes Gustavo Vicentis de Oliveira. Histologic Osseointegration Level Comparing Titanium and Zirconia Dental Implants: Meta-analysis of Preclinical Studies.. The International journal of oral & maxillofacial implants. 2023. DOI: 10.11607/jomi.10142.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.