Investigative radiology · 2024
Multifrequency Magnetic Resonance Elastography Detects Small Abdominal Lymph Node Metastasis by High Stiffness.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Neelsen Christian, Elgeti Thomas, Meyer Tom, Grittner Ulrike, Mödl Lukas, Furth Christian, Geisel Dominik, Hamm Bernd, Sack Ingolf, Marticorena Garcia Stephan Rodrigo
Аннотация
OBJECTIVES: Response Evaluation Criteria in Solid Tumors (RECIST) 1.1 is a clinical and research standard for evaluating malignant tumors and lymph node metastasis. However, quantitative analysis of nodal status is limited to measurement of short axis diameter (SAD), and metastatic lymph nodes below 10 mm in SAD are often not detected. The purpose of this study was to evaluate the value of multifrequency magnetic resonance elastography (MRE) when added to RECIST 1.1 for detection of lymph node metastasis.
MATERIALS AND METHODS: Twenty-five benign and 82 metastatic lymph nodes were prospectively examined by multifrequency MRE at 1.5 T using tomoelastography postprocessing at 30, 40, 50, and 60 Hz (total scan time of 4 minutes). Shear wave speed as a surrogate of soft tissue stiffness was provided in m/s. Positron emission tomography-computed tomography was used as reference standard for identification of abdominal lymph node metastasis from histologically confirmed primary tumors. The diagnostic performance of MRE was compared with that of SAD according to RECIST 1.1 and evaluated by receiver operating characteristic curve analysis using generalized linear mixed models and binary logistic mixed models. Sensitivity, specificity, and predictive values were calculated for different cutoffs.
RESULTS: Metastatic lymph nodes (1.90 ± 0.57 m/s) were stiffer than benign lymph nodes (0.98 ± 0.20 m/s, P < 0.001). An area under the curve of 0.95 for a cutoff of 1.32 m/s was calculated. Using a conservative approach with 1.0 specificity, we found sensitivity (SAD/MRE/MRE + SAD, 0.56/0.84/0.88), negative predictive values (0.41/0.66/0.71), and overall accuracy (0.66/0.88/0.91) to be improved using MRE and even higher for combined MRE and SAD.
CONCLUSIONS: Multifrequency MRE improves metastatic abdominal lymph node detection by 25% based on higher tissue stiffness-even for lymph nodes with an SAD ≤10 mm. Stiffness information is quick to obtain and would be a promising supplement to RECIST.
MATERIALS AND METHODS: Twenty-five benign and 82 metastatic lymph nodes were prospectively examined by multifrequency MRE at 1.5 T using tomoelastography postprocessing at 30, 40, 50, and 60 Hz (total scan time of 4 minutes). Shear wave speed as a surrogate of soft tissue stiffness was provided in m/s. Positron emission tomography-computed tomography was used as reference standard for identification of abdominal lymph node metastasis from histologically confirmed primary tumors. The diagnostic performance of MRE was compared with that of SAD according to RECIST 1.1 and evaluated by receiver operating characteristic curve analysis using generalized linear mixed models and binary logistic mixed models. Sensitivity, specificity, and predictive values were calculated for different cutoffs.
RESULTS: Metastatic lymph nodes (1.90 ± 0.57 m/s) were stiffer than benign lymph nodes (0.98 ± 0.20 m/s, P < 0.001). An area under the curve of 0.95 for a cutoff of 1.32 m/s was calculated. Using a conservative approach with 1.0 specificity, we found sensitivity (SAD/MRE/MRE + SAD, 0.56/0.84/0.88), negative predictive values (0.41/0.66/0.71), and overall accuracy (0.66/0.88/0.91) to be improved using MRE and even higher for combined MRE and SAD.
CONCLUSIONS: Multifrequency MRE improves metastatic abdominal lymph node detection by 25% based on higher tissue stiffness-even for lymph nodes with an SAD ≤10 mm. Stiffness information is quick to obtain and would be a promising supplement to RECIST.