MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/s25154571

Медицинская библиотека · Sensors (Basel, Switzerland) · 2025

Design and Demonstration of a Hybrid FES-BCI-Based Robotic Neurorehabilitation System for Lower Limbs.

Leerskov Kasper S, Spaich Erika G, Jochumsen Mads R, Andreasen Struijk Lotte N S

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

There are only a few available options for early rehabilitation of severely impaired individuals who must remain bedbound, as most exercise paradigms focus on out-of-bed exercises. To enable these individuals to exercise, we developed a novel hybrid rehabilitation system combining a brain-computer interface (BCI), functional electrical stimulation (FES), and a robotic device.

METHODS

The BCI assessed the presence of a movement-related cortical potential (MRCP) and triggered the administration of FES to produce movement of the lower limb. The exercise trajectory was supported by the robotic device. To demonstrate the system, an experiment was conducted in an out-of-lab setting by ten able-bodied participants. During exercise, the performance of the BCI was assessed, and the participants evaluated the system using the NASA Task Load Index, Intrinsic Motivation Inventory, and by answering a few subjective questions.

RESULTS

The BCI reached a true positive rate of 62.6 ± 9.2% and, on average, predicted the movement initiation 595 ± 129 ms prior to the MRCP peak negativity. All questionnaires showed favorable outcomes for the use of the system.

CONCLUSIONS

The developed system was usable by all participants, but its clinical feasibility is uncertain due to the total time required for setting up the system.

Для цитирования: Leerskov Kasper S, Spaich Erika G, Jochumsen Mads R, Andreasen Struijk Lotte N S. Design and Demonstration of a Hybrid FES-BCI-Based Robotic Neurorehabilitation System for Lower Limbs.. Sensors (Basel, Switzerland). 2025. DOI: 10.3390/s25154571.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление