Operator Intracranial Dose Protection During Fluoroscopic-Guided Interventions.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Qazi Emmad, Ursani Ali, Patel Neeral, Kennedy Sean A, Bassett Paul, Jaberi Arash, Rajan Dheeraj, Tan Kong Teng, Mafeld Sebastian
Аннотация
PURPOSE: We utilized an anthropomorphic model made with a human skull to determine how different personal protective equipment influence operator intracranial radiation absorbed dose.
MATERIALS AND METHODS: A custom anthropomorphic phantom made with a human skull coated with polyurethane rubber, mimicking superficial tissues, and was mounted onto a plastic thorax. To simulate scatter, an acrylic plastic scatter phantom was placed onto the fluoroscopic table with a 1.5 mm lead apron on top. Two Radcal radiation detectors were utilized; one inside of the skull and a second outside. Fluoroscopic exposures were performed with and without radiation protective equipment in AP, 45-degree RAO, and 45-degree LAO projections.
RESULTS: The skull and soft tissues reduce intracranial radiation by 76% when compared to radiation outside the skull. LAO (308.95 μSv/min) and RAO projections (96.47μSv/min) result in significantly higher radiation exposure to the primary operator when compared to an AP projection (54 μSv/min). All tested radiation protection equipment demonstrated various reduction in intracranial radiation when compared to no protection. The hood (68% reduction in AP, 91% LAO, and 43% in RAO), full cover (53% reduction in AP, 76% in LAO, and 54% in RAO), and open top with ear coverage (43% reduction in AP, 77% reduction in LAO, and 22% in RAO) demonstrated the most reduction in intracranial radiation when compared to the control.
CONCLUSION: All tested equipment provided various degrees of additional intracranial protection. The skull and soft tissues attenuate a portion of intracranial radiation.
MATERIALS AND METHODS: A custom anthropomorphic phantom made with a human skull coated with polyurethane rubber, mimicking superficial tissues, and was mounted onto a plastic thorax. To simulate scatter, an acrylic plastic scatter phantom was placed onto the fluoroscopic table with a 1.5 mm lead apron on top. Two Radcal radiation detectors were utilized; one inside of the skull and a second outside. Fluoroscopic exposures were performed with and without radiation protective equipment in AP, 45-degree RAO, and 45-degree LAO projections.
RESULTS: The skull and soft tissues reduce intracranial radiation by 76% when compared to radiation outside the skull. LAO (308.95 μSv/min) and RAO projections (96.47μSv/min) result in significantly higher radiation exposure to the primary operator when compared to an AP projection (54 μSv/min). All tested radiation protection equipment demonstrated various reduction in intracranial radiation when compared to no protection. The hood (68% reduction in AP, 91% LAO, and 43% in RAO), full cover (53% reduction in AP, 76% in LAO, and 54% in RAO), and open top with ear coverage (43% reduction in AP, 77% reduction in LAO, and 22% in RAO) demonstrated the most reduction in intracranial radiation when compared to the control.
CONCLUSION: All tested equipment provided various degrees of additional intracranial protection. The skull and soft tissues attenuate a portion of intracranial radiation.