DOI 10.1007/s00270-024-03667-3

Медицинская библиотека · Cardiovascular and interventional radiology · 2024

Discrepancy Between Achieved and Vendor-Predicted Ablation Zones in the Lung: Contributing Factors.

Xu Yiwang, Padley Simon P G, Devaraj Anand, Desai Sujal R, Ridge Carole A

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Several factors are known to affect lung ablation zones. Questions remain as to why there are discrepancies between achieved and vendor-predicted ablation zones and what contributing factors can be modified to balance therapeutic effects with avoidance of complications. This retrospective study of lung tumour microwave ablation analyses day 1 post-treatment CT to assess the effects of lesion-specific and operator-dependent factors on ablation zones.

METHODS AND MATERIALS: Consecutive patients treated at a tertiary centre from 2018 to 2021 were included. All ablations were performed using a single microwave ablation device under lung isolation. The lung tumours were categorised as primary or secondary, and their "resistance" to ablation was graded according to their locations. Intraprocedural pulmonary inflation was assessed as equal to or less than the contralateral non-isolated lung. Ablation energy was categorised as high, medium, or low. Ablation zone dimensions were measured on day 1 CT and compared to vendor reference charts. Ablations with multiple needle positions or indeterminate boundaries were excluded.

RESULTS

A total of 47 lesions in 31 patients were analysed. Achieved long axes are longer than predicted by 5 mm or 14% (p < 0.01) without overall short axis discrepancy. Secondary tumours (p = 0.020), low-resistance location (p < 0.01), good lung inflation (p < 0.01), low (p = 0.003) and medium (p = 0.038) total energy produce lengthened long axes by 4-6 mm or 10-19%. High total energy results in shorter than predicated short axes by 6 mm or 18% (p = 0.010).

CONCLUSION

We identified several factors affecting ablation zone dimensions which may have implications for ablation planning and the avoidance of complications.

Для цитирования: Xu Yiwang, Padley Simon P G, Devaraj Anand, Desai Sujal R, Ridge Carole A. Discrepancy Between Achieved and Vendor-Predicted Ablation Zones in the Lung: Contributing Factors.. Cardiovascular and interventional radiology. 2024. DOI: 10.1007/s00270-024-03667-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.