Split-filter dual energy computed tomography radiotherapy: From calibration to image guidance.

Edmund Jens, Feen Rønjom Marianne, van Overeem Felter Mette, Maare Christian, Margrete Juul Dam Annica, Tsaggari Eirini, Wohlfahrt Patrick

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: Dual-energy computed tomography (DECT) is an emerging technology in radiotherapy (RT). Here, we investigate split-filter DECT throughout the RT treatment chain as compared to single-energy CT (SECT).

MATERIALS AND

METHODS

DECT scans were acquired with a tin-gold split-filter at 140 kV resulting in a low- and high-energy CT reconstruction (recon). Ten cancer patients (four head-and-neck (HN)​, three rectum​, two anal/pelvis and one abdomen) were DECT scanned without and with iodine administered. A cylindrical and an anthropomorphic HN phantom were scanned with DECT and 120 kV SECT. The DECT images generated were: 120 kV SECT-equivalent (CTmix), virtual monoenergetic images (VMIs), iodine map, virtual non-contrast (VNC), effective atomic number (Zeff), and relative electron density (ρe,w). The clinical utility of these recons was investigated for calibration, delineation, dose calculation and image-guided RT (IGRT).

RESULTS

A calibration curve for 75 keV VMI had a root-mean-square-error (RMSE) of 34 HU in closest agreement with the RSME of SECT calibration. This correlated with a phantom-based dosimetric agreement to SECT of γ1%1mm > 98%. A 40 keV VMI recon was most promising to improve tumor delineation accuracy with an average evaluation score of 1.6 corresponding to "partial improvement". The dosimetric impact of iodine was in general < 2%. For this setup, VNC vs. non-contrast CTmix based dose calculations are considered equivalent. SECT- and DECT-based IGRT was in agreement within the setup uncertainty.

CONCLUSIONS

DECT-based RT could be a feasible alternative to SECT providing additional recons to support the different steps of the RT workflow.

Для цитирования: Edmund Jens, Feen Rønjom Marianne, van Overeem Felter Mette, Maare Christian, Margrete Juul Dam Annica, Tsaggari Eirini, Wohlfahrt Patrick. Split-filter dual energy computed tomography radiotherapy: From calibration to image guidance.. Physics and imaging in radiation oncology. 2023. DOI: 10.1016/j.phro.2023.100495.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.