Critical care (London, England) · 2024
Effect of a self-developed fixation device on preventing endotracheal intubation-related pressure injury: a randomised controlled trial.
Рандомизированное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zhang Xiaodong, Zhang Qibing, You Jiqin, Xu Rong, Zhang Zhaojie, Shi Yunlan, Han Chunrong, Zhao Shiyan, Yao Bangtao, Geng Yan, Liu Songqiao
Аннотация
OBJECTIVE: To evaluate the effects of our self-developed endotracheal tube fixation device in mechanically ventilated patients.
METHODS: In a dual-centre randomised controlled trial, patients who were expected to require mechanical ventilation for over 48 h were assigned to the observation group (using self-developed device) or the control group (using the traditional device). The primary endpoint was the incidence of endotracheal intubation-related pressure injury (EIRPI).
RESULTS: Fifty-one patients in the observation group and 54 patients in the control group were analysed. The incidence of EIRPI was 7.8% in the observation group and 33.3% in the control group (p = 0.001). Lip pressure injury (PI) occurred in 0 versus 14 (25.9%) patients in the observation versus control groups (p < 0.001). Both oral-mucosal and facial PIs were similar between the two groups.
CONCLUSIONS: The use of the novel device reduced the incidence of EIRPI, especially lip PI. Trial registration Chinese Clinical Trial Registry ChiCTR2300078132. Registered on 29 November 2023.
METHODS: In a dual-centre randomised controlled trial, patients who were expected to require mechanical ventilation for over 48 h were assigned to the observation group (using self-developed device) or the control group (using the traditional device). The primary endpoint was the incidence of endotracheal intubation-related pressure injury (EIRPI).
RESULTS: Fifty-one patients in the observation group and 54 patients in the control group were analysed. The incidence of EIRPI was 7.8% in the observation group and 33.3% in the control group (p = 0.001). Lip pressure injury (PI) occurred in 0 versus 14 (25.9%) patients in the observation versus control groups (p < 0.001). Both oral-mucosal and facial PIs were similar between the two groups.
CONCLUSIONS: The use of the novel device reduced the incidence of EIRPI, especially lip PI. Trial registration Chinese Clinical Trial Registry ChiCTR2300078132. Registered on 29 November 2023.