Minimizing Pupil Size Dependence in Flicker ERG Using Stiles-Crawford Compensation.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kato Kumiko, Davis C Quentin, Zeinali Nooshin, Sugawara Asako, Matsubara Hisashi, Sugimoto Masahiko, Kashima Yuzen, Mizumoto Keitaro, Kudo Hidetaka, Nagasaka Eiichiro, McCulloch Daphne L, Kondo Mineo
Аннотация
PURPOSE: This study determined the impact of the Stiles-Crawford effect (SCE) on electroretinograms (ERGs). Compensating for the SCE can improve the diagnostic reliability of ERGs by providing a stimulus minimally affected by pupil size.
METHODS: Flicker ERGs were recorded from 10 healthy subjects at 3-minute intervals over 21 minutes after mydriasis. The RETeval system adjusted retinal illuminance in real time based on pupil size measurements, using a Troland stimulus and preset SCE compensation factors (ρdevice) of 0, 0.05, 0.085, and 0.12 mm-2.
RESULTS: Larger pupil areas led to prolonged implicit times with ρdevice = 0 and 0.05 mm-2, whereas ρdevice = 0.12 mm-2 reduced implicit time. Amplitudes were lower with ρdevice = 0 mm-2 but increased with ρdevice = 0.085 and 0.12 mm-2. The values that minimized pupil size dependence were ρdevice = 0.086 mm-2 for the implicit time of the fundamental component of the ERG and ρdevice = 0.05 mm-2 for all other measures. Variability in ERGs based on pupil size is predicted to be ≤7% of the associated 95% reference interval for Troland stimuli over the range of nonmydriatic pupil sizes, compared to ≤43% for luminance stimuli over the range of mydriatic pupil sizes.
CONCLUSIONS: Using Troland stimuli with ρdevice = 0.05 mm-2 for all cone-mediated ERGs would minimize the impact of pupil size, although the improvement would be modest for ERGs performed with Troland stimulation without SCE compensation on non-dilated subjects.
TRANSLATIONAL RELEVANCE: Applying the appropriate SCE coefficient (ρdevice) enables more reliable ERG measurements, improving diagnostic accuracy despite pupil size variations in clinical settings.
METHODS: Flicker ERGs were recorded from 10 healthy subjects at 3-minute intervals over 21 minutes after mydriasis. The RETeval system adjusted retinal illuminance in real time based on pupil size measurements, using a Troland stimulus and preset SCE compensation factors (ρdevice) of 0, 0.05, 0.085, and 0.12 mm-2.
RESULTS: Larger pupil areas led to prolonged implicit times with ρdevice = 0 and 0.05 mm-2, whereas ρdevice = 0.12 mm-2 reduced implicit time. Amplitudes were lower with ρdevice = 0 mm-2 but increased with ρdevice = 0.085 and 0.12 mm-2. The values that minimized pupil size dependence were ρdevice = 0.086 mm-2 for the implicit time of the fundamental component of the ERG and ρdevice = 0.05 mm-2 for all other measures. Variability in ERGs based on pupil size is predicted to be ≤7% of the associated 95% reference interval for Troland stimuli over the range of nonmydriatic pupil sizes, compared to ≤43% for luminance stimuli over the range of mydriatic pupil sizes.
CONCLUSIONS: Using Troland stimuli with ρdevice = 0.05 mm-2 for all cone-mediated ERGs would minimize the impact of pupil size, although the improvement would be modest for ERGs performed with Troland stimulation without SCE compensation on non-dilated subjects.
TRANSLATIONAL RELEVANCE: Applying the appropriate SCE coefficient (ρdevice) enables more reliable ERG measurements, improving diagnostic accuracy despite pupil size variations in clinical settings.