REPEATABILITY OF CHOROIDAL VASCULARITY INDEX MEASUREMENTS USING DIRECTIONAL OPTICAL COHERENCE TOMOGRAPHY IMAGES.

Ersoz Mehmet Giray, Kesim Cem, Karslioglu Melisa Zisan, Yildiz-Tas Ayse, Hasanreisoglu Murat, Sahin Afsun

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the repeatability of choroidal vascularity index (CVI) measurements in centered and decentered (regarding pupil entry position of the beam) directional subfoveal optical coherence tomography (OCT) scans with (CVI1) and without (CVI2) brightness adjustment.

METHODS

Thirty-two eyes of 32 healthy volunteers were included in this prospective study. First, the fovea was evaluated by a single horizontal enhanced depth imaging OCT scan in a centered direction. Then, the same subfoveal horizontal line was scanned in a decentered direction. The agreement between CVIs obtained from these centered and decentered directional OCT scans was investigated using Bland-Altman analysis and intraclass correlation coefficient.

RESULTS

Intraclass correlation coefficient between centered and decentered CVI1 was 0.71 (95% CI: 0.48-0.85, P value < 0.001). Intraclass correlation coefficient between centered and decentered CVI2 was 0.73 (95% CI: 0.5-0.86, P value < 0.001). The mean difference between centered and decentered directional measurements for CVI1 and CVI2 were 0.5 and 0.8, respectively.

CONCLUSION

There is moderate agreement between CVIs obtained from centered and decentered directional single OCT scans of the same subfoveal area. Studies investigating choroidal vascularity should be standardized by using the same OCT beam direction in all scans.

Для цитирования: Ersoz Mehmet Giray, Kesim Cem, Karslioglu Melisa Zisan, Yildiz-Tas Ayse, Hasanreisoglu Murat, Sahin Afsun. REPEATABILITY OF CHOROIDAL VASCULARITY INDEX MEASUREMENTS USING DIRECTIONAL OPTICAL COHERENCE TOMOGRAPHY IMAGES.. Retina (Philadelphia, Pa.). 2021. DOI: 10.1097/iae.0000000000003086.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.