MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/03091902.2020.1742394

Медицинская библиотека · Journal of medical engineering & technology · 2020

Translational development and pre-clinical evaluation of prototype gastrointestinal mock-up devices: only robotic placement of plastic?

Schiemer Jonas F, Merchel Dana, Stumm Karen, Hoffmann Klaus-Peter, Baumgart Jan, Kneist Werner

СтатьяПеревод пока не подготовлен

Аннотация

Background

The aim of this study was to address the vision of wireless theranostic devices distributed along the gastrointestinal (GI) tract by defining design requirements, developing prototype mock-ups, and establishing a minimally invasive surgical approach for the implantation process.

Methods

Questionnaires for contextual analysis and use case scenarios addressing the technical issues of an implantable GI device, a possible scenario for implantation, preparation and calibration of a device, and therapeutic usage by professionals and patients were completed and discussed by an interdisciplinary team of surgeons, engineers, and product designers. Two acute porcine experiments were conducted with a robotic surgical system under general anaesthesia.

Results

A variety of requirements for the design and implantation of implantable devices for modulating GI motility were defined. Five prototype implant mock-ups were three-dimensional (3D)-printed from black polymer material (width 22.32 mm, height 7.66 mm) and successfully implanted on the stomach, duodenum, jejunum, ileum, and colon using the robotic surgical system, without any complications.

Conclusions

Our study shows the development and successful pre-clinical evaluation of a reliable device design with a minimally invasive implantation approach. Several stages of device development, including pre-clinical tests, characterisation of clinical requirements, regulatory affairs, and marketing issues should be managed side by side.

Для цитирования: Schiemer Jonas F, Merchel Dana, Stumm Karen, Hoffmann Klaus-Peter, Baumgart Jan, Kneist Werner. Translational development and pre-clinical evaluation of prototype gastrointestinal mock-up devices: only robotic placement of plastic?. Journal of medical engineering & technology. 2020. DOI: 10.1080/03091902.2020.1742394.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.