Image quality of two different mobile cone beam computed tomographs for maxillofacial surgery.

Seeberger Robin, Buchgeister Markus, Seethaler Andreas C, Shiozawa Thomas, Hoffmann Juergen

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: We evaluated two mobile cone beam computed tomographs (mCBCT) comparing image quality with respect to radiation dosage. Image quality was analyzed by using different scanning modes.

MATERIALS AND

METHODS

The skulls of three human cadavers were scanned by use of conventional Computed Tomography (CT) as well as with two mobile cone beam computed tomographs (Siemens Arcadis Orbic 3D and Ziehm Vision Vario 3D). Six different acquisition modes with different radiation dosages were used. The axial views of all scans were evaluated by five medical doctors regarding image quality by identifying predefined anatomical structures of the skull. A five-point ranking scale was used. The inter-rater reliability was statistically depicted by Cohen's Kappa coefficient. A Wilcoxon signed rank test was used to evaluate the rater's results. For evaluating the signal-to-noise ratio (SNR) a Catphan 600 reference body with two different inlays was used.

RESULTS

Comparing the mCBCTs, the image quality of the Siemens Arcadis Orbic 3D in high-dosage mode received the best score (median: 2.27). The inter-rater reliability was fair (Kappa=-0.030 to 0.328). The Wilcoxon test showed significant (p<0.05) different median rating values in 18 out of 21 imaging modes. The SNR was higher (better) in the high-dosage modes.

CONCLUSION

Intra-operative 3D imaging by using mCBCT for maxillofacial surgery in low-dose mode acquisition is adequate in terms of signal-to-noise ratio and image quality. The image quality does not correlate in a linear manner with a higher radiation dosage. Surgeons using this technique should gather their own experience with the different acquisition modes.

Для цитирования: Seeberger Robin, Buchgeister Markus, Seethaler Andreas C, Shiozawa Thomas, Hoffmann Juergen. Image quality of two different mobile cone beam computed tomographs for maxillofacial surgery.. Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2012. DOI: 10.1016/j.jcms.2012.01.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.