Do biometric parameters improve the quality of optic nerve head measurements with spectral domain optical coherence tomography?

Elksne E, Stingl J V, Schuster A K, Wagner F M, Hoffmann E M

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Spectral domain optical coherence tomography (SD-OCT) is a widely applied non-invasive technique for evaluating optic nerve head parameters. The aim of this study was to evaluate the impact of biometric parameters such as the spherical equivalent (SE) and the anterior corneal curvature (ACC) on the peripapillary retinal nerve fiber layer (pRNFL), Bruch's membrane opening (BMO), and the minimum rim width (MRW) measurements performed by spectral domain optical coherence tomography (SD-OCT) in glaucomatous and healthy eyes.

METHODS

In this cross-sectional, case-control prospective pilot study, the glaucoma group consisted of 50 patients with previously diagnosed and treated glaucoma and one healthy group of 50 subjects. Two consecutive examinations of pRNFL, BMO, and MRW with SD-OCT for every patient were performed without ACC and objective refraction (imaging 1) and with them (imaging 2).

RESULTS

The interclass correlation coefficient (ICC) reflected high agreement between imaging 1 and imaging 2 in both groups. The ICC in the glaucoma and healthy groups for pRNFL (0.99 vs. 0.98), BMO (0.95 vs. 0.97), and MRW (1.0 vs. 1.0) was comparable.

CONCLUSIONS

Our preliminary data from a small number of eyes showed that the measurements of pRNFL, MRW, and BMO reflected high agreement between both imaging techniques with ACC and objective refraction and without these parameters in subjects with a refractive error up to ± 6.0 diopters. Further studies with participants with higher refractive error are necessary to evaluate the impact of biometric parameters such as SE and ACC on measurements with SD-OCT.

Для цитирования: Elksne E, Stingl J V, Schuster A K, Wagner F M, Hoffmann E M. Do biometric parameters improve the quality of optic nerve head measurements with spectral domain optical coherence tomography?. BMC ophthalmology. 2022. DOI: 10.1186/s12886-022-02281-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.