A System for Characterizing Intraoperative Force Distribution During Operative Laryngoscopy.

Ponukumati Aravind S, Wu Xiaotian, Kahng Peter W, Skinner Joseph, Paydarfar Joseph A, Halter Ryan J

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

This study aimed to create and validate an integrated data acquisition system for gauging the force distribution between a laryngoscope and soft-tissue during trans-oral surgery.

METHODS

Sixteen piezoresistive force sensors were interfaced to a laryngoscope and custom maxillary tooth guard. A protocol for calibrating the laryngoscope and maxilla sensors was developed using a motor-controlled linear stage and force measurements were validated against a digital scale. The system was initially tested during suspension laryngoscopy on three cadaver heads mounted on a cadaver head-holder. Intraoperative data was also collected from three patients undergoing head and neck tumor resection.

RESULTS

Mean calibration error of the scope sensors was less than 150 g (n = 3) and mean maxilla sensor error was less than 200 g (n = 3). Peak scope mag-forces of 8.09 ± 6.61 kg and peak maxilla forces of 7.62 ± 4.57 kg were experienced during the cadaver trials. The peak scope sensor mag-force recorded during the intraoperative cases was 24.7 ± 4.53 kg, and the peak maxilla force was 22.0 ± 4.60 kg.

CONCLUSION

The data acquisition system was successfully able to record intraoperative force distribution data. The usefulness of this technology in informing surgeons during trans-oral surgery should be further evaluated in patients with varying anatomic and procedural characteristics.

SIGNIFICANCE: Creation of a low-cost, integrated force-sensing system allows for the characterization of retraction forces at anatomic sites including the pharynx and larynx, brain, and abdomen. Real-time force detection provides surgeons with valuable intraoperative feedback and can be used to improve deformation models at various anatomic sites.

Для цитирования: Ponukumati Aravind S, Wu Xiaotian, Kahng Peter W, Skinner Joseph, Paydarfar Joseph A, Halter Ryan J. A System for Characterizing Intraoperative Force Distribution During Operative Laryngoscopy.. IEEE transactions on bio-medical engineering. 2020. DOI: 10.1109/tbme.2020.2966954.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.