Experimental investigation of aerosol deposition through a realistic respiratory airway replica: An evaluation for MDI and DPI performance.

Ahookhosh Kaveh, Yaqoubi Shadi, Mohammadpourfard Mousa, Hamishehkar Hamed, Aminfar Habib

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: In the present work, a comparison between MDI and DPI for evaluating performance of the devices were carried out by experimentally investigating the deposition parameters through a realistic airway replica.

METHODS

Computed tomography (CT) images of the respiratory airway of a healthy subject were used to develop the realistic model. The airway replica was included extrathoracic, trachea, and tracheobronchial tree up to fourth generations which was fabricated by rapid prototyping. Afterward, in vitro experiments were performed to validate the airway model by comparing the total deposition (G0 to G3) of present replica with available data in the literature. Drug deposition (Salbutamol) in the model was measured by determining concentration of the segments sample by High Performance Liquid Chromatography (HPLC) assay.

RESULTS

Deposition parameters were used for investigating the deposition patterns of the inhaled particles. Results showed that inertial impaction is the dominant mechanism in the most regions of the replica. It was found that the MDI delivered more drug to the tracheobronchial tree compared to the DPI for three different flow rate.

CONCLUSION

The developed realistic respiratory airways model provided an opportunity to more accurately evaluate the performance of drug delivery devices and studying mechanisms of the drug deposition.

Для цитирования: Ahookhosh Kaveh, Yaqoubi Shadi, Mohammadpourfard Mousa, Hamishehkar Hamed, Aminfar Habib. Experimental investigation of aerosol deposition through a realistic respiratory airway replica: An evaluation for MDI and DPI performance.. International journal of pharmaceutics. 2019. DOI: 10.1016/j.ijpharm.2019.05.058.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.