Optimal Depth for Nasopharyngeal Temperature Probe Positioning.
Аннотация
BACKGROUND
The nasopharynx is considered 1 of the 4 generally reliable core temperature measurement sites. But curiously, there is no consensus on how far past the nares to insert the probe. Insertion depth is likely to influence the accuracy of nasopharyngeal temperature measurements because probes near the nares will be cooled by ambient air; similarly, probes inserted too far may approach the airway and be cooled by ventilation gases. We thus determined the range of nasopharyngeal probe insertion depths that best approximate reference core temperature measured in the distal esophagus.
METHODS
In 36 adults undergoing noncardiac surgery with endotracheal intubation, we inserted a nasopharyngeal thermometer 20 cm past the nares and an esophageal temperature probe 40 cm from the incisors. The nasopharyngeal probe was withdrawn sequentially 2 cm at a time at 5-minute intervals. Pairs of nasopharyngeal and reference distal esophageal temperatures were then compared and summarized by Bland and Altman methods.
RESULTS
All nasopharyngeal probe insertion depths between 10 and 20 cm past the nares provided temperatures similar to reference distal esophageal temperatures. At those depths, the bias was typically approximately -0.1°C, with SD of approximately ±0.15°C; the limits of agreement thus were easily within our a priori specified clinically acceptable range of -0.5°C and 0.5°C.
CONCLUSIONS
Any nasopharyngeal probe insertion depth between 10 and 20 cm well represents core temperature in adults having noncardiac surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.