Daily Diagnostic Quality Computed Tomography-on-Rails (CTOR) Image Guidance for Abdominal Stereotactic Body Radiation Therapy (SBRT).
Аннотация
BACKGROUND/
OBJECTIVES
Stereotactic body radiation therapy (SBRT) for abdominal targets faces a variety of challenges, including motion caused by the respiration and digestion and a relatively poor level of contrast between the tumor and the surrounding tissues. Breath-hold treatments with computed tomography-on-rails (CTOR) image guidance is one way of addressing these challenges, allowing for both the tumor and normal tissues to be well-visualized. Using isodose lines (IDLs) from CT simulations as a guide, the anatomical information can be used to shift the alignment or trigger a replan, such that normal tissues receive acceptable doses of radiation.
METHODS
This study aims to describe the workflow involved when using CTOR for pancreas and liver SBRT and demonstrates its effectiveness through several case studies.
RESULTS
In these case studies, using the anatomical information gained through diagnostic-quality CT guidance to make slight adjustments to the alignment, resulted in reductions in the maximum dose to the stomach.
CONCLUSIONS
High-quality imaging, such as CTOR, and the use of IDLs to estimate the doses to OARs, enable the safe delivery of SBRT, without the added complexity and resource commitment required by daily online adaptive planning.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.