DOI 10.1186/1745-6215-13-196

Медицинская библиотека · Trials · 2012

Continuous controllable balloon dilation: a novel approach for cervix dilation.

Arsenijevic Slobodan, Vukcevic-Globarevic Gordana, Volarevic Vladislav, Macuzic Ivan, Todorovic Petar, Tanaskovic Irena, Mijailovic Milan, Raicevic Sasa, Jeremic Branislav

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Cervical dilation using mechanical dilators is associated with various complications, such as uterine perforation, cervical laceration, infections and intraperitoneal hemorrhage. To achieve safe and painless cervical dilation, we constructed a new medical device to achieve confident mechanical cervical dilation: a continuous controllable balloon dilator (CCBD).

METHODS

Controlled pumping of incompressible fluid into the CCBD increases the pressure and outer diameter of the CCBD, continuously dilating the cervical canal. The reliability of the CCBD was confirmed in vitro (testing for consistency and endurance, with no detected risk for breakage) and in vivo. A multi-center clinical study was conducted,with 120 pregnant women randomly assigned to one of three groups: Group I,control group, no dilation;Group II,mechanical dilation, Hegar dilator (HeD); and Group III,CCBD. The tissue material for histological evaluation was obtained from the endocervical mucosa before and after dilation using the HeD or CCBD.

RESULTS

The CCBD dilations were successful and had no complications in all 40 patients of Group III. The cervical tissue was markedly less damaged after CCBD dilation compared with HeD dilation (epithelium damage: 95% (HeD) vs. 45% (CCBD), P <0.001; basal membrane damage: 82.5% (HeD) vs. 27.5% (CCBD), P <0.001; stromal damage: 62.5% (HeD) vs. 37.5% (CCBD), P <0.01). Cervical hemorrhagia was observed in 90% of the patients after HeD dilation versus in 32.5% of the patients after CCBD dilation.

CONCLUSIONS

The CCBD should be used as a replacement for mechanical dilators to prevent uterine and cervical injury during cervical dilation.

TRIAL REGISTRATION: ISRCTN54007498.

Для цитирования: Arsenijevic Slobodan, Vukcevic-Globarevic Gordana, Volarevic Vladislav, Macuzic Ivan, Todorovic Petar, Tanaskovic Irena, Mijailovic Milan, Raicevic Sasa, Jeremic Branislav. Continuous controllable balloon dilation: a novel approach for cervix dilation.. Trials. 2012. DOI: 10.1186/1745-6215-13-196.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.